SECTION 3 GETTING STARTED 3.1 Main Board Power Jumpers

Caution: The meter has no power-on switch, so it will be in operation as soon you apply power.

The meter can be configured to operate on 115VAC or 230VAC by the proper combination of the soldered wire jumpers that are located on the printed circuit board. The meter is set at the factory to be powered by the voltage specified at the time of ordering. The same transformer is used for either configuration, so all you need to do is to select the jumpers as described in this section.

Important: These changes must be performed by a qualified technician.

Note To change the Factory preset jumpers, do the following steps:

Disconnect the power from the unit before proceeding.

1.Remove the main board from the case. Refer to Disassembly/Assembly Section 3.8.

2.Locate the solder jumpers W1, W2, and W3 (located near the edge of the main board alongside the transformer).

3.If your power requirement is 115 V ac, solder jumpers W1 and W3 should be wired, but jumper W2 should not. If your power requirement is 230 Vac, solder jumper W2 should be wired, but jumpers W1 and W3 should not.

5

Image 12
Contents DP119-RTD For immediate technical or application assistance Table of Contents Section Preface Resolution ModelpowerModels Available RTD Models Available Model TypeOther Models Available Page Introduction Unpacking QTY DescriptionSafety Considerations Unpacking & InspectionAbout the Meter Front of the MeterPIN # EXC+EXC Getting Started 3.1 Main Board Power Jumpers AC Voltage Install Remove 1shows the location of solder jumpers W1 through W3Changing the Meter Parameters 2shows the location of the jumper plugs S1 through S5Following tables summarizes what parameters you can change Example of How to Set Meter Option MF1 Jumpers S4 and S5 Sensor Break ProtectionInstallation and Panel Mounting Installation Information Wire RTD Sensor Input ConnectionsWire colors are for Omega RTD wiring only Analog Output Connections Analog Output Connections P22shows the location of the P2 connector on the meter Display Hold ConnectionsMain Power Connections Ð2 shows the location of P1 on the meterDC Wiring 2shows the location of P1 on the meterDisassembly/Assembly Pull TAB Here PC Board Push Here Operation and Calibration Equipment RequiredCalibration Procedure Wire ConfigurationZero Specifications Display Dimensions Glossary Page USAWARRANTY/DISCLAIMER Temperature