Omega DP119-RTD manual Following tables summarizes what parameters you can change

Page 15

The following tables summarizes what parameters you can change.

TABLE 3-1 Range Parameters

 

Option

F/C

Resolution

Range

S1

S4

S5

 

 

*-MF1

¡F

-328 to + 1562¡F

--

A

A

 

 

*-MF2

¡F

0.1¡

-199.9 to + 199.9¡F

C

--

C

 

 

*-MC1

¡C

-200 to + 850¡C

--

A

B

 

 

*-MC2

¡C

0.1¡

-199.9 to + 199.9¡C

C

--

B

 

 

 

 

 

 

 

 

 

 

* Refer to Table 1-1, for complete models (preface section).

TABLE 3-2 2-, 3-, 4-Wire, Upscale, Downscale Configuration

 

Configuration

S2

S3

 

 

S3

 

 

 

 

 

 

Upscale (Default)

 

Downscale

 

 

2- or 4-wire RTD

--

--

 

A, B, E, F

 

C, D, G, H

 

 

3-Wire RTD

A

I

 

A, B, E, F

 

C, D, G, H

 

 

 

 

 

 

 

 

 

 

Note 3-Wire input connection is standard. For 2- or 4-Wire remove S2-A and S3-I

8

Image 15
Contents DP119-RTD For immediate technical or application assistance Table of Contents Section Preface Model Type ResolutionModelpower Models Available RTD Models AvailableOther Models Available Page QTY Description Introduction UnpackingUnpacking & Inspection Safety ConsiderationsFront of the Meter About the MeterPIN # EXC+EXC Getting Started 3.1 Main Board Power Jumpers 1shows the location of solder jumpers W1 through W3 AC Voltage Install Remove2shows the location of the jumper plugs S1 through S5 Changing the Meter ParametersFollowing tables summarizes what parameters you can change Sensor Break Protection Example of How to Set Meter Option MF1 Jumpers S4 and S5Installation and Panel Mounting Installation Information Sensor Input Connections Wire RTDWire colors are for Omega RTD wiring only Analog Output Connections P2 Analog Output ConnectionsDisplay Hold Connections 2shows the location of the P2 connector on the meterÐ2 shows the location of P1 on the meter Main Power Connections2shows the location of P1 on the meter DC WiringDisassembly/Assembly Pull TAB Here PC Board Push Here Equipment Required Operation and CalibrationWire Configuration Calibration ProcedureZero Specifications Display Dimensions Glossary Page USAWARRANTY/DISCLAIMER Temperature