Invacare 636959 instruction sheet Using the Suction Unit, Medical Suction Unit

Page 4

 

Elbow Connector

 

Lid

Holder

Vacuum Gauge

 

 

 

Vacuum

Adjustment

Knob

 

 

Power

Collection

 

 

Bottle

 

 

Connection

Connection

 

Tubing

 

Hydrophobic Filter

Tubing

Power Cord

DETAIL “A”

Fuse

Off (O)

On (I)

Position

Position

FIGURE 1 Medical Suction Unit

Using the Suction Unit

CAUTION

Should fluid be aspirated back into the unit, send your suction unit to an authorized service center for a detailed inspection as the vacuum pump may have been damaged.

NOTE: For this procedure, refer to FIGURE 1 and FIGURE 2 on page 5.

NOTE: If the unit sustains a severe drop, validity of the gauge must be checked. Inspect if the float shut‐off is still functioning properly when the liquid in the collection bottle reaches the safe full level.

NOTE: Always transport unit with vacuum adjust knob rotated fully clockwise in case unit is dropped.

1.Ensure the suction tubing is connected to the labeled patient connector(outlet) on the collection bottle.

2.Ensure that all connections are secure and without leaks.

3.Connect the unit to power.

4.Turn the power switch On (I).

5.Adjust the vacuum level from 0 to 560 mmHg by turning the vacuum adjust knob:

Clockwise to increase vacuum.

Counter‐clockwise to decrease vacuum.

NOTE: Refer to the vacuum gauge while setting the desired level of vacuum. To accurately read the gauge, block the patient end of the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum reading.

Aspirator Suction Unit

4

Part No 636959

Image 4
Contents Safety Summary Aspirator Suction UnitAspirator Suction Unit Grounding Instructions Operating InstructionsSet-Up Using the Suction Unit Medical Suction UnitCleaning the Collection Bottle and Lid Collection Bottle ComponentsCleaning the Tubing Cleaning the Suction UnitMaintenance Hydrophobic Filter Replacement Replacing the Hydrophobic FilterTroubleshooting Replacement Items/AccessoriesLimited Warranty Unité dAspiration Conserver CES InstructionsMise EN Garde Instructions de mise à la terre Consignes d’utilisationMontage Utilisation de lunité daspiration Unité dAspiration MédicalComposants de la bouteille de collecte Nettoyage de la bouteille de collecte et du couvercle Nettoyage de lappareil daspirationNettoyage des tubes Remplacement du filtre hydrophobeEntretien Remplacement du filtre hydrophobe Remplacement des Éléments/Accessoires DépannageGarnatie Limitée Unidad de succión aspiradora Guarde Estas InstruccionesAdvertencia Instrucciones de funcionamiento ConfiguraciónInstrucciones de conexión a tierra Uso de la unidad de succión Figura 1 Unidad de succión médicaLimpieza de la botella de recolección y de la tapa Figura 2 Componentes de la botella de recolecciónLimpieza de la unidad de succión Limpieza de la tuberíaMantenimiento Reemplazo del filtro hidrofóbicoSolución de problemas Problema SoluciónReemplazo de elementos/accesorios Pieza no DescripciónNotas Notas Garantía Limitada USA