Emerson manual Can I calibrate a badly plugged diffuser?, SPS 4000 Troubleshooting

Page 125

Oxymitter 4000

Instruction Manual

IB-106-340 Rev. 3.0 December 2003

3.Always note the time it takes for the cell to recover to the normal process value after the cal gas is removed. As the diffuser plugs, this recovery time will get longer and longer. Use the Calibration Record form provided in this manual.

Can I calibrate a badly plugged diffuser?

1.It may not be possible to immediately replace a plugged diffuser while the process is on line.

2.One can calibrate the probe without pressurizing the cell by adjusting the calibration gas flow rate downward before calibration. For instance, say the process is at 3%, and the first calibration gas is 8%. Adjust the flow of cal gas downward until the reading begins to migrate from 8% to lower values, indicating that process gases are now mixing in with the calibration gases.

3.Adjust the flow rate back up until this mixing is just eliminated. Calibrate at this flow rate. Replace the diffuser at the first opportunity.

Install all protective equipment covers and safety ground leads after trouble- shooting. Failure to replace covers and ground leads could result in seri- ous injury or death.

8-7 SPS 4000 TROUBLESHOOTING

Use the CAL FAIL and IN CAL relay outputs to identify possible SPS 4000 faults.

a.If a calibration was not successfully com- pleted, the SPS 4000 sends a CAL FAIL contact indication to the control room. To determine if the SPS 4000 caused the failed calibration, go to the Oxymitter 4000 site to view the keypad. Or, access the HART/AMS menus. For more information on HART/AMS, refer to Section 7, HART/AMS.

1.If no alarms are indicated on the key- pad or in the HART/AMS STATUS sub-menu, the calibration did not fail because of an Oxymitter 4000 fault. Therefore, a calibration gas flow prob- lem occurred. Refer to Table 8-3or Figure 8-19to troubleshoot the SPS 4000.

2.If the LAST CAL FAILED alarm is indi- cated on the keypad or in the HART/ AMS STATUS sub-menu, the failure is due to either a bad Oxymitter 4000 cell or a calibration gas flow problem.

(a)Verify your calibration setup per paragraph 9-2in Section 9, MAIN- TENANCE AND SERVICE, Also, verify your calibration gas setup.

(b)Perform another calibration and monitor the process. If the calibra- tion fails before both calibration gases finish sequencing, a gas flow problem exists. Refer to Table 8-3or Figure 8-19to troubleshoot the SPS 4000.

If the calibration setup is correct and the Oxymitter 4000 indicates an invalid slope fault (fault 13) be- fore the gases are purged and a last calibration failed fault (fault 15) after the gases are purged, replace the Oxymitter 4000 cell per para- graph 9-4iin Section 9, MAINTE-

NANCE AND SERVICE.

b.If a semi-automatic or manual calibration is being performed but no 5 - 30 VDC relay output contact (IN CAL or CAL FAIL) is be- ing received by the control room, the inter- face board relays are malfunctioning. Replace the interface board per paragraph 9-5b.

NOTE

If the unit is performing frequent auto- calibrations, investigate at the Oxymit- ter 4000 site or using HART/AMS. This condition may indicate an aging cell in the Oxymitter 4000.

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-21

Image 125
Contents IB-106-340 Rev December OxymitterEssential Instructions Summary Highlights of ChangesEffective December, 2003 Rev Highlights of Changes Oxymitter Table of ContentsStartup and Operation of Oxymitter 4000 with LOI Oxymitter 4000 Gas Connections Calibration Gas Connections List of Illustrations11-3 List of Tables11-5 11-1Definitions PrefaceOxymitter Belangrijk Vigtigt Oxymitter Tärkeää Oxymitter Wichtig Importante Viktig Oxymitter Oxymitter Viktigt Oxymitter July 1 Ceramic Fiber Products Material Safety Data SheetSection II. Physical Data Section V. Health Hazard Data Section VI. Reactivity Data Section VII. Spill or Leak Procedures Section IX. Special Precautions Oxymitter Oxymitter 4000 with Remote Imps 4000 Option What YOU Need to KnowOxymitter Remote Electronics with Integral SPS Option Use this Quick Start Guide if Can YOU USE the Following Quick Start GUIDE?Quick Start Guide For Oxymitter 4000 Systems Oxymitter 4000 with SPS 4000 Wiring Diagram Performing a Manual Calibration with a Membrane Keypad Quick Reference Guide Manual Calibration InstructionsTechnical Support Hotline Hart Communicator Fast KEY SequencesSystem Overview Section Description and SpecificationsComponent Checklist of Typical System Package Contents Typical System Package Remote Mounted Oxymitter Integrally MountedMembrane Keypad Model 751 LCD Display Panel Control Room Asset Management Solutions Line Voltage Imps 4000 Multiprobe StandardAutocalibration Option SPS 4000 Single Probe Autocalibration OptionCalibra SPS 4000 Optional Imps 4000 OptionalMounting Components FigureRear View of Manifold only Front ViewDiffusion Elements Model 751 Remote Powered Loop LCD DisplayProbe Options Abrasive Shield Assembly 13. Hastelloy Cup-Type Diffusion AssemblyView a Specifications Oxymitter Nema 4X IP56 100% relative humidityIn. NPT 90 to 250 VAC, 50/60 HzCode Sensing Probe Type Product MatrixPart Number Description Calibration Components3D39695G02 3D39695G013D39695G03 3D39695G04Selecting Location Section InstallationMechanical Installation Oxymitter 4000 Probe Installation Oxymitter 4000 Remote Electronics Installation Vertical Mounted SPS 4000 a Horizontal Mounted SPS 4000 aOxymitter 4000 with Abrasive Shield Oxymitter 4000 Adapter Plate Dimensions Oxymitter 4000 Adapter Plate Installation Oxymitter 4000 Abrasive Shield Bracing Installation Installation with Drip Loop and Insulation Removal Remote Electronics InstallationConnect Line Voltage All wiring must conform to local and national codesIntegral Electronics Install Interconnecting Cable Probe WALL-MOUNTED Electrical Installation for Oxymitter 4000 with SPS Not Used Line Connect 4-20 mA Signal Wiring Connect Relay Output WiringPneumatic Installation for Oxymitter 4000 Without SPS Reference Air Package13. Air Set, Plant Air Connection Imps 4000 Connections Pneumatic Installation forOxymitter 4000 with SPS Oxymitter Verify Mechanical Installation Section Configuration of Oxymitter With Membrane KeypadVerify Terminal Block Wiring GeneralModel 751. The loop-driven LCD display Verify Oxymitter 4000 ConfigurationSW2 Heater T/C Diagnostic Calibration Handshake Signal AlarmLogic I/O Configuration as set at HART/AMS or LOI Mode Logic I/OCalibration Recommended ConfigurationMA Signal Upon Critical Alarm Oxymitter Electronics Housing Terminals and LOI Section Configuration of Oxymitter 4000 with LOIVerify Oxymitter 4000 Configuration Defaults Oxymitter 4000 with LOI Logic I/O Recommended Configuration Oxymitter General Operation Power UPIs an error in the system, one of these Diagnostic Alarm LEDS. If thereTROUBLESHOOTING. Case of multiple errors, only oneO2 0.00% LK warm up 367dgC Section Startup and Operation Oxymitter 4000 with LOIO2 2.59% LK normal Start UP Oxymitter 4000 CalibrationLockout LOI FeaturesData LOI Menu TreeSYSTEM/Calibration Setup Oxymitter 4000 Setup AT the LOISYSTEM/Parameters SYSTEM/Input/OutputSYSTEM/Status LOI InstallationSYSTEM/Software Sensor DataOxymitter 4000 Test Points Model 751 Remote Powered LoopTP3 and TP4 monitor the heater thermo TP1 and TP2 monitor the oxygen cellSignal Line Connections, ≥ 250 Ohms Load Resistance HART/AMS Hart Communicator Signal Line ConnectionsOverview Method 1, For Load Resistance ≥ 250 OhmsLogic I/O Configuration Mode Logic I/O ConfigurationsHart Communicator PC Connections OFF-LINE and ON-LINE OperationsMenu Tree for HART/AMS on the Oxymitter 4000 Sheet 1 Menu Tree for HART/AMS on the Oxymitter 4000 Sheet 2 Menu Tree for HART/AMS on the Oxymitter 4000 Sheet 3 Complete CAL Recommended Apply GAS GAS 1 Flow Hart Communicator Manual O2 CAL MethodDefining a Timed Calibration VIA Hart From the Device Setup screen, selectFrom the Detailed Setup screen, select Oxymitter 100 Section TroubleshootingEMFmV 16.1 18.4 21.1 23.8 27.2 31.2 36.0Alarm Indications Indications with Membrane Keypad Alarm ContactsIdentifying and Correcting Alarm LED Flashes Status MA Line Fault ClearingFault 1, Open Thermocouple Alarms O2 T/C OpenKeypad LOIFault 2, Shorted Thermocouple Alarms O2 T/C ShortedFaulty PC Board Alarms O2 T/C ReversedFault 3, Reversed Thermocouple Wiring or Fault 4, A/D Comm Error Alarms ADC ErrorFault 5, Open Heater Alarms O2 Heater OpenFault 6, High High Heater Temp Alarms Very Hi O2 TempFault 7, High Case Temp Alarms Board Temp HiFault 8, Low Heater Temp Alarms O2 Temp LowFault 9, High Heater Temp Alarms O2 Temp HiFault 10, High Cell mV Alarms O2 Cell OpenFault 11, Bad Cell Alarms O2 Cell BadFault 12, Eeprom Corrupt Alarms EEprom CorruptFault 13, Invalid Slope Fault 14, Invalid Constant Fault 15, Last Calibration Failed Alarms Calib FailedHow do I detect a plugged diffuser? Probe passes calibration, but still appears to read highProbe passes calibration, but still appears to read low Can I calibrate a badly plugged diffuser? SPS 4000 TroubleshootingSPS 4000 Fault Finding Symptom Check Remedy 19. SPS 4000 Troubleshooting Flowchart Sheet 1 Symptom no Test GAS Flow19. SPS 4000 Troubleshooting Flowchart Sheet 2 Calibration Oxymitter 4000 with a Membrane Keypad Section Maintenance and ServiceTion menu Start Calib from the CalibraManual Calibration AlarmsOxymitter CALIBRATION/ Start Calibration Apply Gas Hit E when readyFlow Gas 1xxxxs Read Gas 1xxxxs Done Gas PurgexxxxsReplace Oxymitter 4000 RepairOxymitter 4000 with Integral Electronics, Exploded View Remote Electronics Oxymitter Electronic Assembly Terminal Block Replacement Electronic Assembly ReplacementFuse Replacement Figure Heater Strut Replacement Entire Probe Replacement Excluding Probe HeadHeater Strut Assembly Cell Replacement Probe to Probe Head Assembly Remote Electronics OnlyOxymitter 11. Ceramic Diffusion Element Replacement Oxymitter SPS 4000 Maintenance and Component Replacement Board Replacement 13. SPS 4000 Manifold Assembly Pressure Switch Replacement Solenoid ReplacementInterface Board Power Supply BoardFlowmeter Adjustments Check Valve ReplacementPressure Regulator Optional Maintenance Flowmeter Replacement 15. Calibration Gas and Reference Air Components Value to begin migrating back to the process value Calibration Record For Rosemount Analytical In Situ O2 ProbeSection Return of Material Oxymitter Replacement Parts for Probe Section Replacement PartsFigure and Index No No Dust Seal Description Replacement Parts for Probe Part NumberCell Replacement Kit Oxymitter Probe Disassembly Kit Replacement Parts for Electronics Replacement Parts for Calibration Components Replacement Parts for SPSOxymitter Asset Management Solutions AMS Section Optional AccessoriesHart Handheld 275/375 Communicator BY-PASS PackagesOxymitter Imps 4000 Intelligent Multiprobe Test GAS SequencerSPS 4000 Single Autocalibration Sequencer Catalyst Regeneration O2 Calibration GASOxymitter Section Index Fuse, 8-22, 8-23, 9-18 Warranty Oxymitter Serial no Order no
Related manuals
Manual 32 pages 58.37 Kb

4000 specifications

The Emerson 4000 is a state-of-the-art automation platform designed to enhance industrial processes and improve operational efficiency. This advanced system is recognized for its unparalleled reliability and flexibility, making it suitable for various industries such as oil and gas, chemicals, pharmaceuticals, and power generation.

One of the standout features of the Emerson 4000 is its robust architecture that comprises distributed control systems (DCS) and a comprehensive suite of software applications. This integration allows for real-time monitoring and control of complex processes, ensuring that operators have access to crucial data for informed decision-making. The system supports a wide range of field devices and protocols, facilitating seamless connectivity across various platforms.

The Emerson 4000 employs innovative technologies that elevate its performance. It incorporates advanced analytics and machine learning capabilities, which enable predictive maintenance and reduce downtime. By analyzing historical data and identifying patterns, the system can forecast potential failures, allowing operators to address issues before they escalate. This leads to increased uptime and significant cost savings.

Another key characteristic of the Emerson 4000 is its user-friendly interface. The intuitive design ensures that operators can navigate the system with ease, reducing training time and enhancing productivity. The customizable dashboards provide real-time insights and facilitate quick access to critical information, allowing teams to respond to changes in the process swiftly.

Security is also a cornerstone of the Emerson 4000 platform. The system includes multiple layers of cybersecurity measures to safeguard sensitive data and maintain the integrity of operations. This is crucial in today’s environment, where cyber threats are a significant concern for industrial facilities.

In addition, the Emerson 4000 excels in scalability. It can be easily expanded to accommodate the growing needs of a business without compromising performance. Whether an organization is looking to integrate additional processes or expand its operations geographically, the Emerson 4000 is designed to adapt and grow alongside the business.

Overall, the Emerson 4000 stands out as a powerful tool for industrial automation. Its combination of reliability, advanced technology, user-friendliness, security, and scalability makes it an ideal choice for organizations seeking to optimize their operations and drive efficiency in an increasingly competitive landscape.