Emerson 4000 manual CALIBRATION/ Start Calibration, Apply Gas Hit E when ready, Purgexxxxs

Page 133

Oxymitter 4000

Instruction Manual

IB-106-340 Rev. 3.0 December 2003

9-3 CALIBRATION – OXYMITTER 4000 WITH A LOCAL OPERATOR INTERFACE (LOI)

Refer to Figure 6-4for a view of the LOI menu tree. To calibrate the Oxymitter 4000 from the LOI, access the CALIBRATION/Start Calibration menu.

a.CALIBRATION/ Start Calibration

This is the starting point for calibrations us- ing the LOI. The LOI will instruct the user through this entire procedure. You can se- lect “Abort Calib” at any time to abort the calibration.

1.The LOI displays the following:

Apply Gas 1

Hit E when ready

The Oxymitter 4000 is ready to accept the first calibration gas. Apply the first calibration gas. (Electronics will abort the calibration if this step is not done within 30 minutes).

2.Touch the Enter key to start the Gas 1 flow. A timer is activated to allow the calibration gas adequate time to flow (default time of five minutes). The LOI displays:

Flow Gas 1xxxxs Read Gas 1xxxxs Done Gas 1

The display counts down the seconds remaining to flow Gas 1, then the time remaining for sensing the O2 concen- tration of Gas 1. Done Gas 1 indicates completion.

3.Remove the first calibration gas and apply the second calibration gas. (Electronics will abort the calibration if this step is not done within 30 min- utes). The LOI displays the following:

Apply Gas 2

Hit E when ready

4.Touch the Enter arrow to start the Gas 2 flow. The timer is activated and the LOI displays:

Flow Gas 2xxxxs

Read Gas 2xxxxs

Done Gas 2

Stop Gas

Hit E when ready

5.Remove the second calibration gas and cap off the calibration gas port. Then, touch the Enter arrow to indicate completion. The timer is activated and the LOI displays:

Purgexxxxs

The default purge time is three min- utes. When the gas purge timer times out, the Oxymitter 4000 begins to read the process O2.

b.Abort Calibration

Exits the calibration. After gases are re- moved, and the purge times out, the instru- ment goes back to normal operational mode.

c.Cal Constants – Results of the Calibration

1.Current calibration – If the calibration passed these values will be updated. Log these values onto the calibration log sheet supplied. If the process has high levels of particulate, note the re- sponse back to the process after cal gas is removed, also.

2.Previous Calibration – Values from the prior good calibration

3.Failed Calibration – Bad calibration values are not loaded into the elec- tronics

d.Calibration Status

1.Calibration Step – The current step in an active calibration procedure.

2.Calibration Time – Time until the next scheduled calibration.

3.Next O2 Cal – Time until the next O2 calibration, if different than the next scheduled calibration.

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 9-5

Image 133
Contents IB-106-340 Rev December OxymitterEssential Instructions Effective December, 2003 Rev Highlights of ChangesSummary Highlights of Changes Oxymitter Table of ContentsStartup and Operation of Oxymitter 4000 with LOI Oxymitter 4000 Gas Connections Calibration Gas Connections List of Illustrations11-3 List of Tables11-5 11-1Definitions PrefaceOxymitter Belangrijk Vigtigt Oxymitter Tärkeää Oxymitter Wichtig Importante Viktig Oxymitter Oxymitter Viktigt Oxymitter July 1 Ceramic Fiber Products Material Safety Data SheetSection II. Physical Data Section V. Health Hazard Data Section VI. Reactivity Data Section VII. Spill or Leak Procedures Section IX. Special Precautions Oxymitter Oxymitter 4000 with Remote Imps 4000 Option What YOU Need to KnowOxymitter Remote Electronics with Integral SPS Option Use this Quick Start Guide if Can YOU USE the Following Quick Start GUIDE?Quick Start Guide For Oxymitter 4000 Systems Oxymitter 4000 with SPS 4000 Wiring Diagram Performing a Manual Calibration with a Membrane Keypad Quick Reference Guide Manual Calibration InstructionsTechnical Support Hotline Hart Communicator Fast KEY SequencesComponent Checklist of Typical System Package Contents Section Description and SpecificationsSystem Overview Typical System Package Remote Mounted Oxymitter Integrally MountedMembrane Keypad Model 751 LCD Display Panel Control Room Asset Management Solutions Line Voltage Imps 4000 Multiprobe StandardAutocalibration Option SPS 4000 Single Probe Autocalibration OptionCalibra SPS 4000 Optional Imps 4000 OptionalMounting Components FigureRear View of Manifold only Front ViewProbe Options Model 751 Remote Powered Loop LCD DisplayDiffusion Elements Abrasive Shield Assembly 13. Hastelloy Cup-Type Diffusion AssemblyView a Specifications Oxymitter Nema 4X IP56 100% relative humidityIn. NPT 90 to 250 VAC, 50/60 HzCode Sensing Probe Type Product MatrixPart Number Description Calibration Components3D39695G02 3D39695G013D39695G03 3D39695G04Mechanical Installation Section InstallationSelecting Location Oxymitter 4000 Probe Installation Oxymitter 4000 Remote Electronics Installation Vertical Mounted SPS 4000 a Horizontal Mounted SPS 4000 aOxymitter 4000 with Abrasive Shield Oxymitter 4000 Adapter Plate Dimensions Oxymitter 4000 Adapter Plate Installation Oxymitter 4000 Abrasive Shield Bracing Installation Installation with Drip Loop and Insulation Removal Remote Electronics InstallationConnect Line Voltage All wiring must conform to local and national codesIntegral Electronics Install Interconnecting Cable Probe WALL-MOUNTED Electrical Installation for Oxymitter 4000 with SPS Not Used Line Connect 4-20 mA Signal Wiring Connect Relay Output WiringPneumatic Installation for Oxymitter 4000 Without SPS Reference Air Package13. Air Set, Plant Air Connection Oxymitter 4000 with SPS Pneumatic Installation forImps 4000 Connections Oxymitter Verify Mechanical Installation Section Configuration of Oxymitter With Membrane KeypadVerify Terminal Block Wiring GeneralSW2 Verify Oxymitter 4000 ConfigurationModel 751. The loop-driven LCD display Heater T/C Diagnostic Calibration Handshake Signal AlarmLogic I/O Configuration as set at HART/AMS or LOI Mode Logic I/OMA Signal Upon Critical Alarm Recommended ConfigurationCalibration Oxymitter Electronics Housing Terminals and LOI Section Configuration of Oxymitter 4000 with LOIVerify Oxymitter 4000 Configuration Defaults Oxymitter 4000 with LOI Logic I/O Recommended Configuration Oxymitter General Operation Power UPIs an error in the system, one of these Diagnostic Alarm LEDS. If thereTROUBLESHOOTING. Case of multiple errors, only oneO2 0.00% LK warm up 367dgC Section Startup and Operation Oxymitter 4000 with LOIO2 2.59% LK normal Start UP Oxymitter 4000 CalibrationLockout LOI FeaturesData LOI Menu TreeSYSTEM/Calibration Setup Oxymitter 4000 Setup AT the LOISYSTEM/Parameters SYSTEM/Input/OutputSYSTEM/Status LOI InstallationSYSTEM/Software Sensor DataOxymitter 4000 Test Points Model 751 Remote Powered LoopTP3 and TP4 monitor the heater thermo TP1 and TP2 monitor the oxygen cellSignal Line Connections, ≥ 250 Ohms Load Resistance HART/AMS Hart Communicator Signal Line ConnectionsOverview Method 1, For Load Resistance ≥ 250 OhmsLogic I/O Configuration Mode Logic I/O ConfigurationsHart Communicator PC Connections OFF-LINE and ON-LINE OperationsMenu Tree for HART/AMS on the Oxymitter 4000 Sheet 1 Menu Tree for HART/AMS on the Oxymitter 4000 Sheet 2 Menu Tree for HART/AMS on the Oxymitter 4000 Sheet 3 Complete CAL Recommended Apply GAS GAS 1 Flow Hart Communicator Manual O2 CAL MethodFrom the Detailed Setup screen, select From the Device Setup screen, selectDefining a Timed Calibration VIA Hart Oxymitter 100 Section TroubleshootingEMFmV 16.1 18.4 21.1 23.8 27.2 31.2 36.0Alarm Indications Identifying and Correcting Alarm Alarm ContactsIndications with Membrane Keypad LED Flashes Status MA Line Fault ClearingFault 1, Open Thermocouple Alarms O2 T/C OpenKeypad LOIFault 2, Shorted Thermocouple Alarms O2 T/C ShortedFault 3, Reversed Thermocouple Wiring or Alarms O2 T/C ReversedFaulty PC Board Fault 4, A/D Comm Error Alarms ADC ErrorFault 5, Open Heater Alarms O2 Heater OpenFault 6, High High Heater Temp Alarms Very Hi O2 TempFault 7, High Case Temp Alarms Board Temp HiFault 8, Low Heater Temp Alarms O2 Temp LowFault 9, High Heater Temp Alarms O2 Temp HiFault 10, High Cell mV Alarms O2 Cell OpenFault 11, Bad Cell Alarms O2 Cell BadFault 12, Eeprom Corrupt Alarms EEprom CorruptFault 13, Invalid Slope Fault 14, Invalid Constant Fault 15, Last Calibration Failed Alarms Calib FailedProbe passes calibration, but still appears to read low Probe passes calibration, but still appears to read highHow do I detect a plugged diffuser? Can I calibrate a badly plugged diffuser? SPS 4000 TroubleshootingSPS 4000 Fault Finding Symptom Check Remedy 19. SPS 4000 Troubleshooting Flowchart Sheet 1 Symptom no Test GAS Flow19. SPS 4000 Troubleshooting Flowchart Sheet 2 Calibration Oxymitter 4000 with a Membrane Keypad Section Maintenance and ServiceTion menu Start Calib from the CalibraManual Calibration AlarmsOxymitter CALIBRATION/ Start Calibration Apply Gas Hit E when readyFlow Gas 1xxxxs Read Gas 1xxxxs Done Gas PurgexxxxsReplace Oxymitter 4000 RepairOxymitter 4000 with Integral Electronics, Exploded View Remote Electronics Oxymitter Electronic Assembly Fuse Replacement Figure Electronic Assembly ReplacementTerminal Block Replacement Heater Strut Replacement Entire Probe Replacement Excluding Probe HeadHeater Strut Assembly Cell Replacement Probe to Probe Head Assembly Remote Electronics OnlyOxymitter 11. Ceramic Diffusion Element Replacement Oxymitter SPS 4000 Maintenance and Component Replacement Board Replacement 13. SPS 4000 Manifold Assembly Pressure Switch Replacement Solenoid ReplacementInterface Board Power Supply BoardPressure Regulator Optional Maintenance Check Valve ReplacementFlowmeter Adjustments Flowmeter Replacement 15. Calibration Gas and Reference Air Components Value to begin migrating back to the process value Calibration Record For Rosemount Analytical In Situ O2 ProbeSection Return of Material Oxymitter Replacement Parts for Probe Section Replacement PartsFigure and Index No No Dust Seal Description Replacement Parts for Probe Part NumberCell Replacement Kit Oxymitter Probe Disassembly Kit Replacement Parts for Electronics Replacement Parts for Calibration Components Replacement Parts for SPSOxymitter Asset Management Solutions AMS Section Optional AccessoriesHart Handheld 275/375 Communicator BY-PASS PackagesSPS 4000 Single Autocalibration Sequencer Imps 4000 Intelligent Multiprobe Test GAS SequencerOxymitter Catalyst Regeneration O2 Calibration GASOxymitter Section Index Fuse, 8-22, 8-23, 9-18 Warranty Oxymitter Serial no Order no
Related manuals
Manual 32 pages 58.37 Kb

4000 specifications

The Emerson 4000 is a state-of-the-art automation platform designed to enhance industrial processes and improve operational efficiency. This advanced system is recognized for its unparalleled reliability and flexibility, making it suitable for various industries such as oil and gas, chemicals, pharmaceuticals, and power generation.

One of the standout features of the Emerson 4000 is its robust architecture that comprises distributed control systems (DCS) and a comprehensive suite of software applications. This integration allows for real-time monitoring and control of complex processes, ensuring that operators have access to crucial data for informed decision-making. The system supports a wide range of field devices and protocols, facilitating seamless connectivity across various platforms.

The Emerson 4000 employs innovative technologies that elevate its performance. It incorporates advanced analytics and machine learning capabilities, which enable predictive maintenance and reduce downtime. By analyzing historical data and identifying patterns, the system can forecast potential failures, allowing operators to address issues before they escalate. This leads to increased uptime and significant cost savings.

Another key characteristic of the Emerson 4000 is its user-friendly interface. The intuitive design ensures that operators can navigate the system with ease, reducing training time and enhancing productivity. The customizable dashboards provide real-time insights and facilitate quick access to critical information, allowing teams to respond to changes in the process swiftly.

Security is also a cornerstone of the Emerson 4000 platform. The system includes multiple layers of cybersecurity measures to safeguard sensitive data and maintain the integrity of operations. This is crucial in today’s environment, where cyber threats are a significant concern for industrial facilities.

In addition, the Emerson 4000 excels in scalability. It can be easily expanded to accommodate the growing needs of a business without compromising performance. Whether an organization is looking to integrate additional processes or expand its operations geographically, the Emerson 4000 is designed to adapt and grow alongside the business.

Overall, the Emerson 4000 stands out as a powerful tool for industrial automation. Its combination of reliability, advanced technology, user-friendliness, security, and scalability makes it an ideal choice for organizations seeking to optimize their operations and drive efficiency in an increasingly competitive landscape.