Gentek 3950222, 3940200, 3950320, 3950202, 3940421, 3940320 High Performance Features and Safety

Page 34

Chapter 4: High Performance Features and Safety

Monitor also displays a face velocity value, provides an RS232 output, a night setback mode and several auxiliary relay ports. Both airflow monitors are available options on all XPert models. See Figure 4-1 and 4-2.

11.Inherently Safe Impeller has a negative pressure plenum that surrounds the positive pressure impeller so that if a leak should occur, the unfiltered air is captured and refiltered. See Figure 4-2.

12.Vibration-Isolated Motorized Impeller has vibration isolation supports, which eliminates transfer of vibration to the work surface. This is extremely useful in weighing operations with precision balances. The impeller wheel is also dynamically balanced. See Figure 4-2.

13.Speed Control regulates the speed of the motorized impeller and is used by the certifier to validate and adjust the inflow velocity. Located behind the front panel. See Figure 4-2.

14.Space-Saving Design increases effective laboratory workspace, because the impeller and HEPA filter are contained within the enclosure, a separate filtered blower module is not required. See Figure 4-1

15.Fluorescent Lamp is located above the work area, out of contact with contaminated air. A safety glass window beneath the lamp distributes the light evenly across the work surface. See Figure 4-1.

16.Filter System Pressure Gauge indicates the total system pressure across the HEPA filter and pressurized plenum to help predict HEPA filter loading. See Figure 4-1.

17.Filter Clamping Bolt evenly seals the HEPA filter to the frame of the enclosure. See Figure 4-1 and 4-2.

18.Sash. The sash is constructed of 1/4 inch tempered safety glass. It pivots up for loading and cleaning. See Figure 4-1

19.Control Panel. The control panel, which is located above the sash, contains the filter system pressure gauge, control switches and the electronics. See Figure 4-1.

29

Image 34
Contents User’s Manual USA Warranty Able Ontents PrerequisitesMaintaining Your Filtered Enclosure Chapter Introduction Introduction About This Manual Typographical Conventions Chapter Prerequisites Temperature Variation Requirements Support, Vibration and Movement RequirementsHumidity and Static Electricity Requirements Background on Electrostatics or Static ElectricityExhaust and Blower Requirements Location and Air Current RequirementsSurface Resistivity Table XPert Station Electrical Requirements Space RequirementsChapter Getting Started Lift by the front air foilUnpacking the Enclosure Do not move the enclosure by tilting it onto a hand truckWork Surface Specifications Work Surface and Hepa Filtered Enclosure InstallationFiltered Enclosure Installation XPert Station Getting Started Getting Started Carbon Filters Accessory for Odor Control only Installation of Hepa Filters Accessory Odor Control CarbonHepa Filters Filter Type Appropriate Use 115V ModelsGetting Started Set the Face Velocity with the Speed Control Adjustment Validating the Vented EnclosureSealing the Filtered Enclosure to the Work Surface Chapter High Performance Features and Safety Precautions High Performance FeaturesHigh Performance Features and Safety 1510 Hepa Filtered Enclosure Airflow Diagram High Performance Features and Safety High Performance Features and Safety High Performance Features and Safety High Performance Features and Safety Safety Precautions High Performance Features and Safety High Performance Features and Safety High Performance Features and Safety Routine Daily Work Procedures PlanningWork Techniques Start-upLoading Materials and Equipment Final PurgingHepa Filter Applications Suitability and Guidelines Suitable ApplicationsShutdown Odor Control Carbon Filter Applications sold as an accessory Definition of Terms Appropriate Chemicals for Odor Control Carbon Filters Chemical Carcinogen Use with Odor Control Carbon Filters Prohibited Acid UseBalance Station, which uses a remote blower Chapter Maintaining Your Filtered Enclosure Weekly Routine Maintenance ScheduleMonthly or more often as required Decontamination Determination of when to Replace Hepa FiltersAnnually Maintaining Your Filtered Enclosure Hepa Filter Leak Test PurposeMineral oil Catalog #1491400 Acceptance Calibrate and Operate the Airflow Monitors Labconco Airflow Monitor / Airflow Switch OperationGuardian Airflow Monitor LED Monitor Enclosure Operating In-Flow Alarm Condition Set Point Speed CalibrationGuardian Airflow Monitor LED with Airflow Switch Guardian Digital 1000 Airflow MonitorOperation Digital 1000 CalibrationFactory Settings Guardian Digital 1000 Airflow Monitor Maintaining Your Filtered Enclosure Odor Control Carbon Filter Replacement Procedure See Calculating Odor Control Carbon Filter Life Steps for Calculating PPM and Filter LifeFormaldehyde only Initial CertificationRe-Certification Ammonia onlyFluorescent Light Replacement Motorized Impeller ReplacementMotorized Impeller Replacement Speed Control Replacement Chapter Accessorizing Modifying Your Filtered Enclosure Work SurfacesAirflow Monitor Add appropriate Front Panel with cutout to mount DigitalExhaust Transition Kits for Ducting to Outside Part Description Remote BlowersHepa Filter Bag-In/Bag-Out Bag Accessory Filters Hepa FilterExhaust Dampers Odor Control Carbon FilterStorage Cabinets Utility Shelf KitAccessorizing and Modifying your Filtered Enclosure Chapter Troubleshooting Servicer Operating LOG Problem Cause Corrective ActionTroubleshooting Problem Cause Corrective Action Service Operating Record Log Date CommentsAppendix a Filtered Enclosure Components Replacement Parts Qty Part Number Description Qty Part Number Description Long Use for XPert System and XPert Station Use for XPert System, XPert Station, Digital Monitor and Bag Appendix B Dimensions Exhaust Options Figure B-1 XPert Filtered Balance System Figure B-2 XPert Filtered Balance Station Appendix B Dimensions and Exhaust Options Appendix C Filtered Enclosure Specifications Environmental ConditionsAppendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Ppendix Uick Hart Iltered Nclosures Model SizeNon-mandatory recommendations from Prudent Practices Appendix E References Nfpa 30 2000 Flammable and Combustible Liquids Code General References Laboratory Safety, Principles and Practices, American Page USA