Gentek 3940321, 3940200, 3950320, 3950202, 3940421, 3940320 Maintaining Your Filtered Enclosure

Page 51

Chapter 6: Maintaining Your Filtered Enclosure

How to Install a New HEPA Filter with the Bag-In/Bag-Out Procedure for XPert Filtered Balance Systems and Stations

The XPert enclosures include a true bag-in/bag-out filter disposal system for safely removing the HEPA filter and protect the worker from toxic powders and particulates. See Figures 6-1 for the Bag- In/Bag-Out Procedure. For your safety, wear appropriate personal protective equipment during the change of HEPA filters. If working with biohazardous materials, use the decontamination procedure outlined in Chapter 6. (Consult your Safety Officer before performing filter changes).

1.With the enclosure OFF, remove the front panel secured by two screws on top of the enclosure.

2.Remove the bungee cords and retaining strap on the bag and extend it out. The retaining strap provides an extra level of safety even though the enclosure is under negative pressure when the blower is on.

3.Loosen the two filter clamp bolts with a 9/16" socket securing the HEPA filter. See Figure 4-1 and 4-2.

4.Place both arms in the rear gloves of the bag, and carefully draw the HEPA filter out of the enclosure. Rest the HEPA filter on the floor, table or cart. Remove arms from bag gloves.

5.Create two seals between the enclosure bag access port and the old HEPA filter. Seals can be heat sealed or sealed with tie wraps. Cut off the bag between the two seals and leave the bag stub on the access port to the enclosure.

6.Place the new HEPA filter in the new bag. Then place the new bag and new HEPA filter over the old bag stub and onto the enclosure access port.

7.Carefully remove the old bag stub using the single glove near the access port. Reattach the retaining strap on the perimeter of the new bag at the enclosure access port.

Draw the old bag stub into the single glove and turn the glove inside out. Double seal this glove and cut off the glove with the contaminated bag stub inside.

8.Carefully install the new HEPA filter and uniformly tighten the two filter clamp bolts until the HEPA filter gasket is compressed 50%.

46

Image 51
Contents User’s Manual USA Warranty Prerequisites Able OntentsMaintaining Your Filtered Enclosure Chapter Introduction Introduction About This Manual Typographical Conventions Chapter Prerequisites Support, Vibration and Movement Requirements Temperature Variation RequirementsBackground on Electrostatics or Static Electricity Humidity and Static Electricity RequirementsLocation and Air Current Requirements Exhaust and Blower RequirementsSurface Resistivity Table XPert Station Space Requirements Electrical RequirementsLift by the front air foil Chapter Getting StartedDo not move the enclosure by tilting it onto a hand truck Unpacking the EnclosureWork Surface and Hepa Filtered Enclosure Installation Work Surface SpecificationsFiltered Enclosure Installation XPert Station Getting Started Getting Started Installation of Hepa Filters Accessory Odor Control Carbon Carbon Filters Accessory for Odor Control onlyHepa Filters 115V Models Filter Type Appropriate UseGetting Started Validating the Vented Enclosure Set the Face Velocity with the Speed Control AdjustmentSealing the Filtered Enclosure to the Work Surface High Performance Features Chapter High Performance Features and Safety PrecautionsHigh Performance Features and Safety 1510 Hepa Filtered Enclosure Airflow Diagram High Performance Features and Safety High Performance Features and Safety High Performance Features and Safety High Performance Features and Safety Safety Precautions High Performance Features and Safety High Performance Features and Safety High Performance Features and Safety Planning Routine Daily Work ProceduresFinal Purging Start-upLoading Materials and Equipment Work TechniquesSuitable Applications Hepa Filter Applications Suitability and GuidelinesShutdown Odor Control Carbon Filter Applications sold as an accessory Definition of Terms Appropriate Chemicals for Odor Control Carbon Filters Prohibited Acid Use Chemical Carcinogen Use with Odor Control Carbon FiltersBalance Station, which uses a remote blower Chapter Maintaining Your Filtered Enclosure Routine Maintenance Schedule WeeklyMonthly or more often as required Determination of when to Replace Hepa Filters DecontaminationAnnually Maintaining Your Filtered Enclosure Purpose Hepa Filter Leak TestMineral oil Catalog #1491400 Acceptance Labconco Airflow Monitor / Airflow Switch Operation Calibrate and Operate the Airflow MonitorsGuardian Airflow Monitor LED Monitor Calibration Enclosure Operating In-Flow Alarm Condition Set Point SpeedGuardian Digital 1000 Airflow Monitor Guardian Airflow Monitor LED with Airflow SwitchDigital 1000 Calibration OperationFactory Settings Guardian Digital 1000 Airflow Monitor Maintaining Your Filtered Enclosure Odor Control Carbon Filter Replacement Procedure See Steps for Calculating PPM and Filter Life Calculating Odor Control Carbon Filter LifeAmmonia only Initial CertificationRe-Certification Formaldehyde onlyMotorized Impeller Replacement Fluorescent Light ReplacementMotorized Impeller Replacement Speed Control Replacement Work Surfaces Chapter Accessorizing Modifying Your Filtered EnclosureAdd appropriate Front Panel with cutout to mount Digital Airflow MonitorExhaust Transition Kits for Ducting to Outside Remote Blowers Part DescriptionOdor Control Carbon Filter Accessory Filters Hepa FilterExhaust Dampers Hepa Filter Bag-In/Bag-Out BagUtility Shelf Kit Storage CabinetsAccessorizing and Modifying your Filtered Enclosure Problem Cause Corrective Action Chapter Troubleshooting Servicer Operating LOGTroubleshooting Problem Cause Corrective Action Date Comments Service Operating Record LogAppendix a Filtered Enclosure Components Replacement Parts Qty Part Number Description Qty Part Number Description Long Use for XPert System and XPert Station Use for XPert System, XPert Station, Digital Monitor and Bag Appendix B Dimensions Exhaust Options Figure B-1 XPert Filtered Balance System Figure B-2 XPert Filtered Balance Station Appendix B Dimensions and Exhaust Options Environmental Conditions Appendix C Filtered Enclosure SpecificationsAppendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Appendix C Filtered Enclosure Specifications Model Size Ppendix Uick Hart Iltered NclosuresNon-mandatory recommendations from Prudent Practices Appendix E References Nfpa 30 2000 Flammable and Combustible Liquids Code General References Laboratory Safety, Principles and Practices, American Page USA