Excell Precision FM user manual Internal Function Parameter Description, Function Setting

Page 20

EXCELL PRECISION CO., LTD.

2Internal Function Parameter Description

Parameter

Function

 

Setting

Default

Code

Parameter

Description

Setting

 

CFN-00

CFN-01

CFN-02

CFN-03

CFN-04

CFN-05

Tare or Zero

0

ON

0

Function under

1

OFF

Unstable Status

 

 

 

 

Re-zeroing after

0

OFF

0

Turning on

1

ON

 

 

 

 

 

 

 

 

 

0%: Full range re-zero

 

Re-zero Range

0% ~ 30%

1% ~ 30%:

0

 

 

Capacity × ± setting value%

 

Filter Strength

0 ~ 5

Strength increases by number

2

 

 

 

 

 

0

No limit

 

AD Sampling

1

20 times/sec.

2

Rate

2

10 times/sec.

 

 

3

5 times/sec.

 

 

0

OFF

 

Animal Scale

1

Mode 1: No weight display

0

under unstable status

Mode

 

 

 

 

2Mode 2: Weight display whether under stable or unstable status

CFN-06

Animal Scale

0 ~ 100

Mode 2: Stable Range Setting

30

Stable Range

 

 

 

 

 

 

0

8 times

 

 

Animal Scale

1

16 times

 

CFN-07

Sampling

2

32 times

2

 

Frequency

3

64 times

 

 

 

 

 

 

4

128 times

 

CFN-08

Dual Range

0

Multi-interval

0

Resolution

1

Multi-range

 

Setting

 

 

Dual Range

 

 

 

CFN-09

Resolution

0 ~ 65535

 

5000

Middle Point

 

 

 

 

 

Setting

FM User Manual

19

SME300000119

Image 20
Contents SME300000119 Table of Content Transmission Interface Safety Precaution Before Using the IndicatorFeatures and Specification FeaturesSpecification Analog SpecificationDigital Specification Power RequirementKeypad Operation Instruction Function Operation DescriptionFront & Rear Panels Front PanelIndication KeypadRear Panel Load Cell Chapter InstallationWired 5-wired Load Cell Dimension Battery Assembly Internal Calibration Specification Setting Parameter CodeDescription Specification Parameter DescriptionSetting Display Decimal Point Digit None Weight Calibration WeightPassword Setting Internal Function Setting Parameter Description Function SettingInternal Function Parameter Description Error Message Animal Scale Function Setting Special FunctionCFN-08 = 1 ⇒ Multi-range Dual Range Resolution Switch FunctionCFN-08 = 0 ⇒ Multi-interval Resolution Switch Function Pre-tare CancellationFNC-02 or FNC-03 setting is at parameter 5. HR Pre-tare FunctionPin Allocation of Rear Panel Transmission InterfaceOP-01 RS232/RS485 Serial Output with RTC Connection Description RS485 Function Parameter Setting To escape Function Parameter Setting RS232 Setting Press Enter Input the *Parameter code intendedOP-01 RS232 / RS485 Interface Function Parameter Function CodeNo Limit Times/sec Transmission Format CR LFCommand Mode Command FunctionMaintenance Default Recovery for All ParametersDefault Recovery for General Function Parameters Self-diagnosis Mode2 7-segment Display Testing Keypad & Calibration Switch TestingEeprom Testing RTC Time & Date Testing