Excell Precision FM user manual Weight Calibration

Page 17

EXCELL PRECISION CO., LTD.

5-2 Internal Weight Calibration

Turn on and warm up the machine for 15 to 30 minutes before calibration.

Adjust calibration switch to “ON”, and the screen will display

.

Press TARE or ZERO to select

.

 

Procedure

ENTER

Zero

Calibration

ESC

Weight

Calibration

ESC

Calibration

Switch “OFF”

ENTER

Zero calibration complete

ENTER

Place the

ENTER

 

weight mass

Confirmed

 

and set the

 

 

 

weight

 

Zero Calibration

a ) Ensure nothing on the platter; after being stable, press ENTER , and the screen will display

……. ”. The zero calibration will be complete 5 seconds later.

b ) To abort zero calibration, just press

.

ESC

Weight Calibration

a ) Place an object, whose weight is known, on the platter, and input the weight value from

front panel. After being stable, press

ENTER and the screen will display “……. ”. The

weight calibration will be complete 5 seconds later.

b ) To abort weight calibration, just press

.

ESC

FM User Manual

16

SME300000119

Image 17
Contents SME300000119 Table of Content Transmission Interface Before Using the Indicator Safety PrecautionFeatures Features and SpecificationAnalog Specification SpecificationDigital Specification Power RequirementFunction Operation Description Keypad Operation InstructionFront Panel Front & Rear PanelsIndication KeypadRear Panel Load Cell Chapter InstallationWired 5-wired Load Cell Dimension Battery Assembly Internal Calibration Parameter Code Specification SettingDescription Specification Parameter DescriptionSetting Display Decimal Point Digit None Weight Weight CalibrationPassword Setting Internal Function Setting Parameter Description Function SettingInternal Function Parameter Description Error Message Special Function Animal Scale Function SettingCFN-08 = 1 ⇒ Multi-range Dual Range Resolution Switch FunctionCFN-08 = 0 ⇒ Multi-interval Pre-tare Cancellation Resolution Switch FunctionFNC-02 or FNC-03 setting is at parameter 5. HR Pre-tare FunctionPin Allocation of Rear Panel Transmission InterfaceOP-01 RS232/RS485 Serial Output with RTC Connection Description RS485 Function Parameter Setting To escape Function Parameter Setting Press Enter Input the *Parameter code intended RS232 SettingParameter Function Code OP-01 RS232 / RS485 Interface FunctionNo Limit Times/sec CR LF Transmission FormatCommand Function Command ModeDefault Recovery for All Parameters MaintenanceDefault Recovery for General Function Parameters Self-diagnosis ModeKeypad & Calibration Switch Testing 2 7-segment Display TestingEeprom Testing RTC Time & Date Testing