Excell Precision FM user manual No Limit Times/sec

Page 32

EXCELL PRECISION CO., LTD.

 

 

 

0

No Limit

 

 

 

 

1

1 times/sec.

 

 

RS1-08

Transmission

2

2 times/sec

4

 

Times

3

5 times/sec

 

 

 

 

 

 

4

10 times/sec

 

 

 

 

5

20 times/sec

 

 

RS1-09

Date Setting

 

 

 

 

 

 

 

 

 

 

RS1-10

Time Setting

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FM User Manual

31

SME300000119

Image 32
Contents SME300000119 Table of Content Transmission Interface Safety Precaution Before Using the IndicatorFeatures and Specification FeaturesSpecification Analog SpecificationDigital Specification Power RequirementKeypad Operation Instruction Function Operation DescriptionFront & Rear Panels Front PanelIndication KeypadRear Panel Load Cell Chapter InstallationWired 5-wired Load Cell Dimension Battery Assembly Internal Calibration Specification Setting Parameter CodeDescription Specification Parameter DescriptionSetting Display Decimal Point Digit None Weight Calibration WeightPassword Setting Internal Function Setting Parameter Description Function SettingInternal Function Parameter Description Error Message Animal Scale Function Setting Special FunctionCFN-08 = 1 ⇒ Multi-range Dual Range Resolution Switch FunctionCFN-08 = 0 ⇒ Multi-interval Resolution Switch Function Pre-tare CancellationFNC-02 or FNC-03 setting is at parameter 5. HR Pre-tare FunctionPin Allocation of Rear Panel Transmission InterfaceOP-01 RS232/RS485 Serial Output with RTC Connection Description RS485 Function Parameter Setting To escape Function Parameter Setting RS232 Setting Press Enter Input the *Parameter code intendedOP-01 RS232 / RS485 Interface Function Parameter Function CodeNo Limit Times/sec Transmission Format CR LFCommand Mode Command FunctionMaintenance Default Recovery for All ParametersDefault Recovery for General Function Parameters Self-diagnosis Mode2 7-segment Display Testing Keypad & Calibration Switch TestingEeprom Testing RTC Time & Date Testing