TROUBLESHOOTING
PROBLEM | PROBABLE CAUSE | CORRECTIVE MEASURES |
1) Display shows "LOBAT" | Batteries are low | a) Replace batteries (Refer to page 3). |
| a) Bad connection between Light Wand | a) Be sure the Light Wand is properly |
2) Display shows "1999.9" or | and Probe Socket | inserted into the Probe Socket. |
|
| |
| b) Too much/too little light | b) Be sure the Light Wand is properly |
|
| seated in the Optical Standard when |
|
| Zero Adjusting. |
3) Bad Repeatability | a) Inconsistent measurement technique | a) Make sure you are applying enough |
|
| pressure between the Light Wand and |
|
| the subject's arm. |
|
| b) The silver line on the Light Wand |
|
| must be aligned with the line on the |
|
| Optical Standard when Zero Adjusting. |
|
| c) When taking a measurement, the |
|
| silver line on the Light Wand must point |
|
| toward the subject's shoulder. |
|
| d) Use the Light Shield to block the |
|
| external light. Always avoid |
|
| measurements under strong lights (e.g. |
|
| direct sunlight) because errors may |
|
| occur. |
29