Smeg WO-01 manual List of faults

Page 16

6- Faults

Front door leds

The door has two leds to indicate:

-Green: power on, normal operation;

-Red: fault, production of osmosis-treated water stopped

Important:

THE DOOR GIVES NO INDICATION OF DELIVERY

CIRCUIT FAULTS

See below for details.

Green

Red

List of faults

A) No osmosis-treated water delivery

Possible causes

Subcause

Solution

Display alarm

message

 

 

 

 

 

 

 

 

 

 

No water in vat

 

See point "B"

see point "B"

 

 

 

 

 

 

 

 

Check power supply and power switch ON

 

 

No electrical power

 

(see “Positioning and installation”)

 

 

 

Power plug not inserted in electrical panel

 

 

 

 

 

 

 

 

(insert)

 

 

 

 

Check connections and whether board

 

 

No power to pump

 

is sending a signal

 

Delivery pump not

 

 

Pump condenser faulty (replace)

 

 

 

Empty vat (manually with jumper on delivery

 

running

Vat pressure switch faulty (jammed on

 

 

 

 

pump or cock –fig.4) and replace pressure

 

 

empty)

 

 

 

 

switch (fig.1)

 

 

 

 

 

 

 

 

 

 

Delivery pressure switch faulty (jammed

 

Replace, or adjust the pressure with the

 

 

on full)

 

adjuster screw (fig.2)

 

 

Non-return valve jammed (does not

 

Check valve, if jammed replace/clean (fig.3)

none

 

open)

 

 

 

 

 

Vat not full enough for use (min level)

 

Wait for a few minutes while it fills

 

 

 

 

 

 

 

No water arriving at pump: vat suction

 

Remove obstacle/open cock (fig.4)

 

 

pipe blocked/crimped/cock closed

 

 

 

 

 

 

 

 

 

 

 

Delivery pump runs

Air in delivery pipe

 

Shake pipe and see if air rises into the vat

 

 

 

but does not deliver

Impeller damaged/broken

 

Replace

 

or only insufficiently

 

 

 

 

The user requires minimum intake

 

System limit

 

 

pressure of >1.8 BAR

 

 

 

 

 

 

 

The user requires flow > 20 l/min

 

Reduce user load

 

 

 

 

 

 

- 42 -

Image 16
Contents WO-01 Page Contents General warnings Technical characteristics 150-600 KPa Bar Electrical hookup PositioningPlumbing hookup Page Pressurizing and starting up the unit Delivery pump startup first startupReplacing the pre-filter Replacing the membranesSpecial adjustments InactivityPlumbing diagram FIL Wiring diagram Hook UP J2 Green Power LED RES List of faults Osmosis-treated water not produced Water production not conforming Page Electronic circuit board and programming Page