6- Faults
Front door leds
The door has two leds to indicate:
-Green: power on, normal operation;
-Red: fault, production of
Important:
THE DOOR GIVES NO INDICATION OF DELIVERY
CIRCUIT FAULTS
See below for details.
Green
Red
List of faults
A) No osmosis-treated water delivery
Possible causes | Subcause | Solution | Display alarm | ||
message | |||||
|
|
|
| ||
|
|
|
|
| |
| No water in vat |
| See point "B" | see point "B" | |
|
|
|
|
| |
|
|
| Check power supply and power switch ON |
| |
| No electrical power |
| (see “Positioning and installation”) |
| |
|
| Power plug not inserted in electrical panel |
| ||
|
|
|
| ||
|
|
| (insert) |
| |
|
|
| Check connections and whether board |
| |
| No power to pump |
| is sending a signal |
| |
Delivery pump not |
|
| Pump condenser faulty (replace) |
| |
|
| Empty vat (manually with jumper on delivery |
| ||
running | Vat pressure switch faulty (jammed on |
|
| ||
|
| pump or cock |
| ||
| empty) |
|
| ||
|
| switch (fig.1) |
| ||
|
|
|
| ||
|
|
|
| ||
| Delivery pressure switch faulty (jammed |
| Replace, or adjust the pressure with the |
| |
| on full) |
| adjuster screw (fig.2) |
| |
|
| Check valve, if jammed replace/clean (fig.3) | none | ||
| open) |
| |||
|
|
| |||
| Vat not full enough for use (min level) |
| Wait for a few minutes while it fills |
| |
|
|
|
|
| |
| No water arriving at pump: vat suction |
| Remove obstacle/open cock (fig.4) |
| |
| pipe blocked/crimped/cock closed |
|
| ||
|
|
|
| ||
|
|
|
|
| |
Delivery pump runs | Air in delivery pipe |
| Shake pipe and see if air rises into the vat |
| |
|
| ||||
but does not deliver | Impeller damaged/broken |
| Replace |
| |
or only insufficiently |
|
|
|
| |
The user requires minimum intake |
| System limit |
| ||
| pressure of >1.8 BAR |
|
| ||
|
|
|
| ||
| The user requires flow > 20 l/min |
| Reduce user load |
| |
|
|
|
|
|
- 42 -