11.The previous calibrated operation of the crosslinker will continue following any unsuccessful recalibration. Replace the new UV wavelength tubes with the previously removed old wavelength tubes.

12.Successful recalibration to the NEW UV wavelength changes the values in the microprocessor and numeric setting value is displayed on the LED. To return to original or another wavelength the proper UV calibration sensor is needed.

- 9 -

Image 11
Contents Ultraviolet Crosslinkers Contents Introduction Important Safety InformationDescription and Specifications CL-1000L Longwave Crosslinker FeaturesElectrical CL-1000 Shortwave CrosslinkerOperation UV Crosslinkers UVP UV Crosslinkers UVP Applications MAINTENANCE, Care and Cleaning of the Crosslinker Changing the UV Wavelength UV Wavelength Calibration ProcedureUV Crosslinkers UVP Crosslinker Calibration Procedure Equipment necessary Voltage Range for U4 Range for U5 Operations with TOP Cover Removed Replacement Parts and Servicing ProceduresOperations With Top Cover Removed Removal of top cover General Servicing ProceduresReplacement of tube holders Replacement of starter holder Replacement of startersReplacement of main switch Replacement of fuse holdersReplacement of main connector Replacement of membrane panel Replacement of UV sensorReplacement of PCB Replacement of reflector Changing plastic feet General Items Changing fusesSchematic Wiring Diagram Corporate Headquarters UVP, Inc