UVPS 81-0112-01 Rev C manual Replacement of reflector

Page 19

UV CROSSLINKERS

UVP

C.Operations With The Internal Chassis Assembly Removed Removal of the assembly

1.Release the wire to the micro switch located on the bottom left internal chassis assembly.

2.Release the main power cable to ballasts - black wire with connector (color red).

3.Release all eight nuts positioned four on top of the unit, two sets on either side of the unit. See diagram for positions.

4.Slide unit out from the main frame.

Replacement of reflector

The reflector is located inside of the internal chassis assembly.

1.Release the eight screws-four on the top and two on either side of the unit.

2.Pull the reflector out towards the front of the unit. Note: for cleaning the reflector, refer to Section 6 for maintenance/care/cleaning of the crosslinker.

3.For replacement-reverse the above procedure.

D.Operations With The Reflector Removed Replacement of ballasts

The ballasts are fitted to the right hand side of the internal chassis assembly.

1.Select the ballast(s) which need to be replaced.

2.Cut the tie wraps to locate the ballast electrical connectors.

3.Cut these wires and mark positions.

4.Re-fit new ballasts using M4 posidrive screws with shake proof and plain washers with a nut as fitted on the diagram.

-17 -

Image 19
Contents Ultraviolet Crosslinkers Contents Description and Specifications Important Safety InformationIntroduction CL-1000L Longwave Crosslinker FeaturesElectrical CL-1000 Shortwave CrosslinkerOperation UV Crosslinkers UVP UV Crosslinkers UVP Applications MAINTENANCE, Care and Cleaning of the Crosslinker Changing the UV Wavelength UV Wavelength Calibration ProcedureUV Crosslinkers UVP Crosslinker Calibration Procedure Equipment necessary Voltage Range for U4 Range for U5 Operations with TOP Cover Removed Replacement Parts and Servicing ProceduresReplacement of tube holders General Servicing ProceduresOperations With Top Cover Removed Removal of top cover Replacement of starter holder Replacement of startersReplacement of main connector Replacement of fuse holdersReplacement of main switch Replacement of PCB Replacement of UV sensorReplacement of membrane panel Replacement of reflector Changing plastic feet General Items Changing fusesSchematic Wiring Diagram Corporate Headquarters UVP, Inc