UVPS 81-0112-01 Rev C manual Voltage Range for U4 Range for U5

Page 13

UV CROSSLINKERS

UVP

9.Run the Crosslinker in calibration mode by depressing the STOP/RESET button and turning the power switch on at the same time. At this point, both pens should start recording information on paper. Sometimes either pen will record zero information regardless of the lights in the crosslinker being on, if this is the case, check for bad connections.

10.During calibration, check the voltage levels on both pens. By adjusting the potentiometers on the PBC, adjust the values for U4 and U5. Depending on the voltage used, from past units it has been found that the optimum values for these levels are:

VOLTAGE RANGE FOR U4

RANGE FOR U5

115V

3.3-3.5V

2.1-2.3V

220V

3.5-3.7V

1.7-2.0V

100V

2.5-2.7V

2.6-2.9V

11.When the calibration cycle is finished, move both pens to UP and ZERO positions.

12.Unplug the UVX-25 sensor and disconnect the probe for PEN 2.

13.At this point, it is necessary to check the energy levels of the crosslinker to see if the unit was calibrated properly. Slide the UVX-25's cable through the lower hole of the inner chassis and plug it into the radiometer.

14.Connect the radiometer to PEN 2 of the chart recorder.

15.Select 1V scale for PEN 2 and a speed of 60mm/min.

16.Lower both pens, turn the CHART DRIVE switch on, change the pens' switches to MEASURE position and adjust for zero level.

17.Select an energy level of 600 Joules on the crosslinker and press the START button.

18.When the cycle is completed, repeat for 1,200 and 2,000 Joules levels.

19.Calculate the actual amount of energy and compare it with the energy selected. They should be within a range of +/- 10%. If this is not the case, adjustments should be made during a repeat calibration. Repeat all the above procedures until the actual levels are within range.

20.When the three levels of energy are acceptable, disconnect the sensor and the chart recorder from the Crosslinker. At this point, the unit is ready to be tested.

-11 -

Image 13
Contents Ultraviolet Crosslinkers Contents Description and Specifications Important Safety InformationIntroduction Electrical FeaturesCL-1000 Shortwave Crosslinker CL-1000L Longwave CrosslinkerOperation UV Crosslinkers UVP UV Crosslinkers UVP Applications MAINTENANCE, Care and Cleaning of the Crosslinker Changing the UV Wavelength UV Wavelength Calibration ProcedureUV Crosslinkers UVP Crosslinker Calibration Procedure Equipment necessary Voltage Range for U4 Range for U5 Operations with TOP Cover Removed Replacement Parts and Servicing ProceduresReplacement of tube holders General Servicing ProceduresOperations With Top Cover Removed Removal of top cover Replacement of starter holder Replacement of startersReplacement of main connector Replacement of fuse holdersReplacement of main switch Replacement of PCB Replacement of UV sensorReplacement of membrane panel Replacement of reflector Changing plastic feet General Items Changing fusesSchematic Wiring Diagram Corporate Headquarters UVP, Inc