7.CHANGING THE UV WAVELENGTH
A UV Crosslinker is purchased with shortwave, longwave or midrange tubes. However, if user requirements and applications change, the Crosslinkers provide you with the unique ability to change the ultraviolet wavelength and recalibrate the UV sensor and microprocessor to the new UV wavelength. This is accomplished by purchasing FIVE tubes of the NEW UV wavelength and the proper UVP calibration sensor.
TUBE P/NWAVELENGTH | UVP CALIBRATION SENSOR P/N | ||
365 nm Longwave | |||
302 nm Midrange | |||
254 nm Shortwave |
UV Wavelength Calibration Procedure
1.After disconnecting the unit from the electrical supply, install the FIVE NEW WAVELENGTH Tubes as directed in Section 9A. (Changing tubes).
2.Plug the corresponding UV calibration sensor into the TOP HOLE just inside of the door at the upper right.
3.Place the UV calibration sensor in the middle of the floor of the exposure chamber.
4.Close the crosslinker door.
5.Invoke the calibration mode of the Crosslinker by pushing/holding down STOP on the tactile touchpad and turning the power to the unit ON.
6.A sequence of tones will be heard from the Crosslinker and a 180 second count will display on the LED when the calibration procedure has started.
7.No calibration occurs during this
8.Upon completion of the
9.If sensor readings are unacceptable, an error code (01, 02, 03 or 04) will flash on the LED Press any key to stop.
-8 -
10.Check all connections and redo calibration. If the same error message appears, call UVP.