Fluke 189 user manual Testing for Continuity, Automatically subtract this value

Page 44

Model 187 & 189

Users Manual

Keep the following in mind when measuring resistance:

Because the meter’s test current flows through all possible paths between the probe tips, the measured value of a resistor in a circuit is often different from the resistor’s rated value.

The test leads can add 0.1 to 0.2 of error to

 

resistance measurements. To test the leads, touch

 

the probe tips together and read the resistance of the

 

leads. If necessary, you can press D to

 

automatically subtract this value.

The resistance function can produce enough voltage to forward-bias silicon diode or transistor junctions, causing them to conduct. To avoid this, do not use the 30 Mor 500 Mranges for in-circuit resistance measurements.

Testing for Continuity

Caution

To avoid possible damage to the meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before testing for continuity.

Continuity is the presence of a complete path for current flow. The continuity test features a beeper that sounds if a circuit is complete. The beeper allows you to perform quick continuity tests without having to watch the display.

The continuity function detects intermittent opens and shorts lasting as little as 1 millisecond (0.001 second). These brief contacts cause the meter to emit a short beep.

To select continuity, turn the rotary switch to resistance position, then press the blue button once. The continuity symbol (~) appears in the display. Continuity uses manual ranging only; autoranging is not available. Refer to Figure 3-6 for continuity testing setup instructions.

3-8

Image 44
Contents Model 187 Lifetime Limited Warranty Table of Contents Model 187 Making MeasurementsIii Maintenance SpecificationsList of Tables Model 187 List of Figures Viii Contacting Fluke Safety InformationBefore You Start Safety Information WWarningBefore You Start Symbols International Electrical SymbolsTurning the Meter On Getting AcquaintedIntroduction Battery Considerations Automatic Power OffLow Battery Indication Automatic Backlight OffRotary Switch Rotary SwitchPushbuttons PushbuttonsRotary Switch Selections Position Rotary Switch Function Blue Key FunctionRotary Switch Positions View MEMButton Pushbuttons Yellow Button Description Function Pushbuttons Button Description Yellow Button FunctionPrimary Display Selecting the RangeUnderstanding the Display Bar Graph Secondary DisplayDisplay Features Number Feature Display Features Description Display Features Number 8.8.8Number Feature Display Features DescriptionDisplay Features Number 51000 FMS8888 HM8888 8.8Using the Input Terminals Input TerminalsUsing Display Hold Display Hold and AutoHOLDUsing AutoHOLD Using MIN MAXModel 187 Using Fast MN MX Min Max AvgUsing Relative Mode REL Using Hold with MIN MAX or Fast MN MXMaking Measurements Measuring VoltageMeasuring AC Voltage AC Voltage MeasurementDB Measurements in AC Volts Functions DBm DisplayMeasuring DC Voltage Both AC and DC Voltage Measurements AC over DC DC over AC AC and DC DisplayMeasuring Resistance Capacitors before measuring resistanceOFF Testing for Continuity Automatically subtract this valueUsing Conductance for High Resistance Tests For in-circuit tests, turn circuit power off Conductance Measurement Measuring Capacitance Testing Diodes Capacitance MeasurementDiode Test Electrically live circuits Measuring TemperatureShock, do not connect the thermocouples to Measuring Current Input RangesInput Alert Feature Measuring CurrentMeasuring AC Current Following are some tips for measuring current11. AC Current Measurement Measuring DC Current Total current to circuit Current through one componentMeasuring Frequency 13. Functions Allowing Frequency MeasurementMeasuring Duty Cycle 14. Hz DisplayMeasures Positive Pulse Measures Negative PulseMeasuring Pulse Width 16. Duty Cycle DisplayStart and Stop on Positive Slope Start and Stop on Negative Slope18. Pulse Width Display Model 187 Using Memory & Communications Features Types of MemorySaved Readings Memory Logged Readings MemoryStopping Logging Storing Saved ReadingsStarting Logging Viewing Memory Data To avoid electric shock, disconnect the testView Display MEMUsing Communications 187 Clearing MemoryModel 187 Changing the Default Settings Selecting Setup OptionsFunction Specific Setup Selections Factory DefaultPrecondition OptionCommon Setup Selections ChoicesAdjusting the Temperature Offset Changing the Default Settings Selecting Display Resolution 3 1/2 or 4 1/2 Digits Setting the Power Off TimeoutSetting the 24-Hour Clock Setting the Line Main FrequencyReturning to Factory Defaults Saving Setup OptionsTesting the Fuses MaintenanceIntroduction General Maintenance Replace F2 if Replace F1 if Displayed Replacing the Batteries Battery and Fuse Replacement Case of Difficulty Replacing the FusesUser-Replaceable Parts MP8 MP9 Specifications Safety and CompliancesPhysical Specifications Feature Summary Basic Specifications Detailed Accuracy Specifications Accuracy Function 45-1 kHz20 kHz KHz-100 kHzDC a 500.00 Ω Capacitance Diode Test MIN MAX AVG Frequency Counter Sensitivity Burden Voltage A, mA, µAInput Characteristics