2.8 Koehler illumination

Koehler illumination needs to be adjusted in the beginning of every imaging session for optimal illumination. Proceed as follows:

Instruction

View in eyepiece

1.Use 10x objective and front lens. Focus on the sample.

2.Close the luminous-field diaphragm by turning the adjustment wheel clockwise.

3.Move the condenser up or down so that the diaphragm edges are in focus. The microscope is designed so that in the imaging path, field diaphragm should be in conjugate plane with the specimen plane (in focus at the same time). You adjust this by adjusting the height of the condenser.

6

Image 8
Contents 2008 Saving and Transferring Images Settings for transmitted light microscopy General user informationTurning on the microscope XY-stagePlacing the sample ObjectivesFocusing Auxiliary condenser lens switchEye pieces oculars Dual viewing attachmentDual viewing attachment Koehler illumination InstructionInstruction Polarization contrast FiltersPage Accessing the computer View Finder application Adjustments before image acquisitionCapture resolution Image acquisitionWhite balance Exposure ModeLevels adjustment Averaging Focus toolOver/Underexposure Post-imaging adjustments Capturing the imageViewing images Cropping the imageScale bars Saving and transferring imagesFinishing your imaging Troubleshooting Computer login fails