simply multiplied values, thus the same effect can be added later on in image editing software. Therefore, changing the value from 100 is not recommended.

3.3.3Exposure Mode

In the Exposure Mode box, select either Automatic or Manual:

Automatic (AE): The software measures the optimal exposure time automatically. Change Spot and Exposure Adjust if needed (see below). To lock the exposure time, click AE lock button (Figure 15, upper toolbar, rightmost button).

oSpot defines the size of an area from which exposure time is measured. Choose 30 %, 1 % or 0.1 % of picture area. To move the spot, there is a button in the upper toolbar (Figure 15, upper toolbar, third button from right). Click the button and then click live view image in a place where you want the exposure area to be. Re-centering the area is done with the button on the right-hand side of the Spot button.

oExposure Adjust is used for adjusting the automatic exposure time. Use values from 0 to 2 to brighten (longer exposure) the image and values from -2to 0 to darken (shorter exposure) the image.

Manual: You can adjust the exposure time manually using the Exposure Time slider.

3.3.4White balance

White balance is the adjustment of relative intensities in red, green, and blue channels. White balance adjustment is needed because the halogen lamp in the microscope changes color with the change of intensity. The dimmer the light, the more red it becomes.

Pressing the button marked with OWB will automatically adjust the white point of the image. If manual adjustment is needed, press the WB-button and make a selection in the image that represents white color. The image is adjusted accordingly. The best white balance is achieved by pressing the OWB-button while the sample is out of the view (the whole camera view area is white) so it does not disturb the measurement of the color of the light.

3.3.5Levels adjustment

You can adjust the brightness, contrast, and gamma correction of the image in the levels adjustment dialog (Main Menu/BF-button/Settings/Level Adjustment tab). Start the adjustment by clicking the levels button (marked with red square in Figure 16). The window contains an image histogram, which shows the intensity distribution of the image. Wide histogram indicates that good dynamic range of the camera was used. A peak at the right end means that there are oversaturated pixels in the image. You should avoid over saturation by shortening the exposure time.

The Channel selection lets you adjust one color at a time. In the RGB setting, your adjustments will affect all colors. Adjust Shadow and Highlight for brightness and contrast, and Gamma for gamma correction. Note that Gamma is a non-linear change and has to be mentioned in the legend when the image is published.

11

Image 13
Contents 2008 Saving and Transferring Images General user information Settings for transmitted light microscopyTurning on the microscope XY-stageObjectives Placing the sampleAuxiliary condenser lens switch FocusingDual viewing attachment Eye pieces ocularsDual viewing attachment Instruction Koehler illuminationInstruction Filters Polarization contrastPage Adjustments before image acquisition Accessing the computer View Finder applicationCapture resolution Image acquisitionLevels adjustment Exposure ModeWhite balance Over/Underexposure Focus toolAveraging Capturing the image Post-imaging adjustmentsViewing images Cropping the imageFinishing your imaging Saving and transferring imagesScale bars Troubleshooting Computer login fails