Emerson CMF200A Maximum wiring distances, Maximum cable lengths, Pipe run, Valves

Page 11

Determining a Location

Maximum wiring distances

If the transmitter is mounted remotely from the sensor, the maximum distance between the sensor and transmitter depends on cable type. See Table 2.

For high-temperature and extreme high-temperature ELITE sensors, note the following:

For sensors with a Model 1700/2700 or a Model 2400S transmitter, the transmitter is considered to be integrally mounted on the sensor, so Table 2 does not apply.

For sensors with a junction box or core processor, the limits in Table 2 apply only to the wiring between the junction box or core processor and a remotely-mounted transmitter. The length of the flexible conduit on which the junction box or core processor is mounted does not need to be considered.

Table 2 Maximum cable lengths

Cable type

Wire size

Maximum length

 

 

 

Micro Motion 9-wire to an MVD

Not applicable

60 feet (20 meters)

transmitter or core processor

 

 

 

 

 

Micro Motion 9-wire to all other

Not applicable

1000 feet (300 meters)

transmitters

 

 

 

 

 

Micro Motion 4-wire

Not applicable

1000 feet (300 meters)

 

 

 

User-supplied 4-wire

 

 

Power wires (VDC)

22 AWG (0,35 mm2)

300 feet (90 meters)

 

 

 

 

20 AWG (0,5 mm2)

500 feet (150 meters)

 

 

 

 

18 AWG (0,8 mm2)

1000 feet (300 meters)

 

 

 

Signal wires (RS-485)

22 AWG (0,35 mm2) or larger

1000 feet (300 meters)

Pipe run

Micro Motion sensors do not require a straight run of pipe upstream or downstream.

Valves

Meter zero can be adjusted for special circumstances. When zeroing is performed with the auto-zero feature of the transmitter, it is necessary to halt flow through the sensor, with the sensor tubes full of process fluid at typical process conditions. The valve used to halt flow in this situation should be downstream from the sensor. It is also possible to restore the meter’s factory zero setting. For more information about zeroing, refer to the instruction manual shipped with the transmitter.

Installation Manual

9

Image 11
Contents Micro Motion Elite Sensors Page Before You Begin Customer serviceTroubleshooting Installation options Specifications DefinitionsEuropean installations CMF025, CMF050, or CMF100 Elite sensor with core processor Elite sensor with junction box Environmental limits Hazardous area installationsDetermining a Location Full flow tubes113 Sensor models Electronics UL and CSA ambient temperature limitsValves Maximum wiring distancesMaximum cable lengths Pipe runVertical pipeline Orienting the SensorRecommended sensor orientations Flow direction arrowMounting an Elite sensor Mounting the SensorFeedthrough, plastic cap, and clamping ring Installing extended mount electronicsMounting the temperature extender onto the sensor Installing wafer-style sensors Wafer-style assembly Core processor to a 4-wire remote transmitter WiringWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Elite pressure ratings GroundingPressure Ratings Removing a purge plug Case purging procedurePurge Fittings Time required to purge Elite sensor cases Rupture DisksUsed equipment Return PolicyNew and unused equipment 20002158