Emerson CMF200A installation manual Grounding, Pressure Ratings, Elite pressure ratings

Page 22

Grounding

Step 5 Grounding

The sensor can be grounded via the piping if the joints in the pipeline are ground-bonded. If the sensor is not grounded via the piping, connect a ground wire to the internal or external grounding screw, which is located on the transmitter, core processor, or junction box.

Improper grounding can cause measurement error. Ground the flowmeter to earth, or follow ground network requirements for the facility.

If national standards are not in effect, follow these guidelines to ground the sensor:

Use copper wire, 14 AWG (2,0 mm²) or larger wire size for grounding.

Keep all ground leads as short as possible, less than 1 ohm impedance.

Connect ground leads directly to earth, or follow plant standards.

Pressure Ratings

Table 4 shows the pressure ratings for ELITE sensors.

Table 4 ELITE pressure ratings

 

 

EN-1092 flange

 

 

 

 

Pressure rating (bar)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Note

PN

Type

Facing

Description

ASTM

50°C

100°C

150°C

200°C

250°C

300°C

350°C

400°C

427°C

 

 

 

 

material

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

40

11

B1

Flange, EN

304

40.0

34.4

30.8

 

28.0

26.0

24.1

 

23.0

22.0

21.4

1

 

weld

 

1092-1 PN40

 

 

 

 

 

 

 

 

 

 

 

 

 

 

neck

 

Form B1 (304)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

40

11

B2 & D

Flange, EN

316

40.0

40.0

36.3

 

33.7

31.8

29.7

 

28.5

27.4

26.9

2

 

weld

 

1092-1 PN40

 

 

 

 

 

 

 

 

 

 

 

 

 

 

neck

 

Form B2 & D

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(316)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100

11

B2 & D

Flange, EN

316

100.0

100.0

90.9

 

84.2

79.5

74.2

 

71.4

68.5

67.3

2

 

weld

 

1092-1 PN100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

neck

 

Form B2 & D

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(316)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

40

32/02

B1

Flange, EN

C-22 collar/

40.0

34.4

30.8

 

28.0

26.0

24.1

 

23.0

22.0

21.4

3

 

lap

 

1092-1 PN40

304 flange

 

 

 

 

 

 

 

 

 

 

 

 

 

joint

 

Form B1(lap

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

joint)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.Pressure rating limited by flange (for 304 sensor tubes).

2.Pressure rating limited by flange (for 316 sensor tubes).

3.Pressure rating limited by flange (for C-22 Alloy sensor tubes).

20

Micro Motion® ELITE® Sensors

Image 22
Contents Micro Motion Elite Sensors Page Troubleshooting Customer serviceBefore You Begin European installations Specifications DefinitionsInstallation options CMF025, CMF050, or CMF100 Elite sensor with core processor Elite sensor with junction box Full flow tubes Hazardous area installationsDetermining a Location Environmental limits113 UL and CSA ambient temperature limits Sensor models ElectronicsPipe run Maximum wiring distancesMaximum cable lengths ValvesFlow direction arrow Orienting the SensorRecommended sensor orientations Vertical pipelineMounting the Sensor Mounting an Elite sensorInstalling extended mount electronics Feedthrough, plastic cap, and clamping ringMounting the temperature extender onto the sensor Installing wafer-style sensors Wafer-style assembly Wiring Core processor to a 4-wire remote transmitterWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Pressure Ratings GroundingElite pressure ratings Purge Fittings Case purging procedureRemoving a purge plug Rupture Disks Time required to purge Elite sensor casesNew and unused equipment Return PolicyUsed equipment 20002158