Page 14

Mounting the Sensor

Installing extended mount electronics

If you ordered a sensor with an extended mount for the electronics, you will need to install the extender. To install the extended mount electronics:

1.There is a metal clamping ring around the base of the feedthrough (Figure 9). Remove this ring and set it aside. You will use it later to secure the extender to the sensor.

2.The sensor was shipped with a plastic cap over the feedthrough pins. Remove the plastic cap and discard it.

3.There is a plastic plug inside the extended mount (Figure 10). Remove the plug and discard it.

4.Place the extended mount onto the feedthrough and carefully rotate the extender until the feedthrough notches line up. Being careful not to damage the feedthrough pins, push the extender down onto the feedthrough until it is fully engaged.

5.Replace the clamping ring on the feedthrough. Tighten the ring’s screw to 13–18 in-lbs (1.5–2 N-m).

Figure 9 Feedthrough, plastic cap, and clamping ring

Plastic cap — remove and discard

Clamping ring — remove but do not discard

Clamping ring screw

Feedthrough

12

Micro Motion® ELITE® Sensors

Image 14
Contents Micro Motion Elite Sensors Page Before You Begin Customer serviceTroubleshooting Installation options Specifications DefinitionsEuropean installations CMF025, CMF050, or CMF100 Elite sensor with core processor Elite sensor with junction box Full flow tubes Hazardous area installationsDetermining a Location Environmental limits113 UL and CSA ambient temperature limits Sensor models ElectronicsPipe run Maximum wiring distancesMaximum cable lengths ValvesFlow direction arrow Orienting the SensorRecommended sensor orientations Vertical pipelineMounting the Sensor Mounting an Elite sensorInstalling extended mount electronics Feedthrough, plastic cap, and clamping ringMounting the temperature extender onto the sensor Installing wafer-style sensors Wafer-style assembly Wiring Core processor to a 4-wire remote transmitterWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Elite pressure ratings GroundingPressure Ratings Removing a purge plug Case purging procedurePurge Fittings Rupture Disks Time required to purge Elite sensor casesUsed equipment Return PolicyNew and unused equipment 20002158