Emerson CMF200A installation manual 113

Page 9

Determining a Location

Figure 5 ELITE sensor ambient and process temperature limits (all models except high-temperature and extreme high-temperature models)

Mount transmitter remotely; use junction box

(°C)

176 (80)

 

 

 

 

140

 

 

 

 

°F

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(60)

 

 

 

 

in

140 (60)

 

 

 

 

 

 

 

 

 

transmitter

104 (40)

 

 

 

 

 

 

 

 

113 (45)

 

 

 

 

 

 

 

 

 

68 (20)

 

 

 

 

 

 

 

 

 

or

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

processor

32 (0)

 

 

 

 

 

 

 

 

 

–4 (–20)

 

 

 

 

 

 

 

 

 

core

–40 (–40)

 

 

 

 

 

 

 

 

 

of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

temperature

–76 (–60)

 

 

 

 

 

 

 

 

 

 

 

Mount transmitter remotely; use junction box

 

 

 

–112 (–80)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ambient

–148 (–100)

 

 

 

 

 

 

 

 

 

–400 (–240)

–292 (–180)

–184 (–120)

–76 (–60)

32 (0)

140 (60)

248 (120)

356 (180)

400 (204)

464 (240)

Maximum process temperature in °F (°C)

Notes:

1.When ambient temperature is below –40 °F (–40 °C), a core processor or Model 2400S transmitter must be heated to bring its local ambient temperature to between –40 °F (–40 °C) and +140 °F (+60 °C). Long-term storage of electronics at ambient temperatures below –40 °F (–40 °C) is not recommended.

2.Temperature limits may be further restricted by hazardous area approvals.

3.The extended mount option allows the sensor case to be insulated without covering the transmitter, core processor, or junction box, but does not affect temperature ratings.

4.For the purposes of selecting electronics options, this graph should be used only as a general guide. If your process conditions are close to the gray areas, it may be inappropriate to use electronics options other than a junction box. Consult with your Micro Motion representative.

Installation Manual

7

Image 9
Contents Micro Motion Elite Sensors Page Customer service TroubleshootingBefore You Begin Specifications Definitions European installationsInstallation options CMF025, CMF050, or CMF100 Elite sensor with core processor Elite sensor with junction box Determining a Location Hazardous area installationsFull flow tubes Environmental limits113 Sensor models Electronics UL and CSA ambient temperature limitsMaximum cable lengths Maximum wiring distancesPipe run ValvesRecommended sensor orientations Orienting the SensorFlow direction arrow Vertical pipelineMounting an Elite sensor Mounting the SensorFeedthrough, plastic cap, and clamping ring Installing extended mount electronicsMounting the temperature extender onto the sensor Installing wafer-style sensors Wafer-style assembly Core processor to a 4-wire remote transmitter WiringWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Grounding Pressure RatingsElite pressure ratings Case purging procedure Purge FittingsRemoving a purge plug Time required to purge Elite sensor cases Rupture DisksReturn Policy New and unused equipmentUsed equipment 20002158