Emerson CMF200A Determining a Location, Full flow tubes, Hazardous area installations

Page 8

Determining a Location

Figure 4 High-temperature or extreme high-temperature ELITE sensor

Flow direction arrow

Process connection

Transmitter, core processor, or junction box

(Model 2400S transmitter shown)

CMF200A, B, C, or E CMF300A, B, C, or E CMF400A or C CMFHC3A or CMFHC3C

Purge connections (optional)

Flexible conduit

Approval tag

Calibration tag

Step 1 Determining a Location

Choose a location for the sensor based on the requirements described in this section. The following general guidelines can help you select an appropriate location for the sensor.

Full flow tubes

For optimal performance, the sensor tubes should remain full of process fluid.

Hazardous area installations

Make sure the hazardous area specified on the sensor approval tag is suitable for the environment in which the sensor is installed. (See Figures 1–4.) For installation in an area that requires intrinsic safety, refer to the appropriate Micro Motion approval documentation, shipped with the sensor or available from the

Micro Motion web site at www.micromotion.com.

Improper installation in a hazardous area can cause an explosion. When installing in a hazardous area, refer to Micro Motion approvals instructions, shipped with the product or available from the

Micro Motion web site. For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.

Environmental limits

The ambient and process temperature limits of the sensor are shown in Figures 5 and 6:

For ELITE sensor models other than high-temperature and extreme-high temperature models, see Figure 5.

For high-temperature and extreme high-temperature ELITE sensors, see Figure 6.

6

Micro Motion® ELITE® Sensors

Image 8
Contents Micro Motion Elite Sensors Page Before You Begin Customer serviceTroubleshooting Installation options Specifications DefinitionsEuropean installations CMF025, CMF050, or CMF100 Elite sensor with core processor Elite sensor with junction box Hazardous area installations Determining a LocationFull flow tubes Environmental limits113 UL and CSA ambient temperature limits Sensor models ElectronicsMaximum wiring distances Maximum cable lengthsPipe run ValvesOrienting the Sensor Recommended sensor orientationsFlow direction arrow Vertical pipelineMounting the Sensor Mounting an Elite sensorInstalling extended mount electronics Feedthrough, plastic cap, and clamping ringMounting the temperature extender onto the sensor Installing wafer-style sensors Wafer-style assembly Wiring Core processor to a 4-wire remote transmitterWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Elite pressure ratings GroundingPressure Ratings Removing a purge plug Case purging procedurePurge Fittings Rupture Disks Time required to purge Elite sensor casesUsed equipment Return PolicyNew and unused equipment 20002158