Measurement Accuracy and Speed
Measurement Accuracy and Speed
The power meter has no internal ranges. The only ranges you can set are those of the HP
Caution | To prevent damage to your sensor do not exceed the power levels |
| specified in the section “Maximum Power” on page 35. |
| The Agilent 9304A sensor is DC coupled. DC voltages in excess of the |
| maximum value (5 Vdc) can damage the sensing diode. |
|
|
Setting the Range
There are two manual settings, “LOWER” and “UPPER”. The LOWER range uses the more sensitive path and the UPPER range uses the attenuated path in the HP
Table 1 Sensor Ranges
Sensor | LOWER range | UPPER range |
|
|
|
E9300/1/4A | ||
|
|
|
E9300/1B | +20 dBm to +44 dBm | |
|
|
|
E9300/1H | 0 dBm to +30 dBm | |
|
|
|
The default is “AUTO”. In AUTO the range crossover value depends on the sensor model being used (see Table 2).
Table 2 Range Crossover Values
E9300/1/4A | E9300/1B | E9300/1H |
|
|
|
+20 dBm ± 0.5 dBm | 0 dBm ± 0.5 dBm | |
|
|
|
28