E9300/1/4/A Power Sensor Specifications
Switching Point
The Agilent
+20 dBm. The power meter automatically selects the proper power level path. To avoid unnecessary switching when the power level is near the
| Error |
|
|
Offset at Switch Point | ≤± 0.5% (≤± 0.02 dB) typical |
|
|
Switching Point Hysteresis
0.5 dB typical
Zero Set,
Zero Drift and
Measurement Noise
| Conditions | Zero Set | Zero Drift2 | Measurement | |
| (RH)1 | Noise3 | |||
|
|
| |||
|
|
|
|
| |
Lower Range | 15% to 75% | 500 pW | 150 pW | 700 pW | |
75% to 95% | 500 pW | 4,000 pW | 700 pW | ||
10 dBm) |
|
|
|
| |
Upper Range | 15% to 75% | 500 nW | 150 nW | 500 nW | |
75% to 95% | 500 nW | 3,000 nW | 500 nW | ||
+20 dBm) |
|
|
|
| |
|
|
|
|
|
1.RH is the abbreviation for Relative Humidity.
2.Within 1 hour after zero set, at a constant temperature, after a 24 hour
3.The number of averages at 16 for Normal mode and 32 for x2 mode, at a constant temperature, measured over a one minute interval and two standard deviations.
39