TO ADJUST THE FREQUENCY:

Look at the meter. The kHz indicator must blink to be able to tune the frequency.

If it is not blinking, “Click” in the Sensitivity control dial one time and the kHz indicator in the display panel will blink.

When the kHz indicator blinks, change the frequency by rotating the Sensitivity dial up (clockwise) or down (counter clockwise).

YELLOW STORE BUTTON:

TO STORE A READING:

“click” / press the yellow Store Button. This puts the instrument in the data storage mode. In the data storage mode the display panel changes.

The Storage Location is shown in the upper left corner. There are 400 Storage Locations numbered 001 to 400. If the Storage Location has no data in it, the display will show: “NOT USED”.

If there has been information stored in the selected location, the upper section of the display will indicate that information. The text field (if previously selected), Time, Date, Decibel, Frequency and Operation Mode “R”, “S”, “P” (RO, SO, or PO with offset Value in the Offset Mode) will blink and alternate (scroll). The text field, if previously selected in the Set Up Mode, may be used to record notes or codes.

The lower left corner of the display indicates the current decibel level selected for storage.

The lower right of the display indicates the current frequency selected for storage.

The lower right corner of the display indicates the Operation Mode “R”, “S” or “P”, RO, SO, or PO.

001

NOT USED

25dB

40KHz R

 

 

Data Storage Mode Display

TO STORE DATA:

Click the Store Button again and the data will be stored and displayed on top.

TO OVERWRITE DATA OR TO ENTER DATA IN A NEW LOCATION:

Press the yellow Store button to enter the data storage mode.

Spin the Sensitivity Dial until the desired storage location is displayed on the screen

Click the yellow Store Button to store the new information in that location and proceed as described above.

NOTE: When using Ultratrend software it is possible to enter a new reading that is out of sequence by spinning to the last unused memory location (as long as all 400 locations are not filled) and entering the data as described above. Following instructions in Ultratrend, a new sequential order can be updated to include the new item(s) for future inspections.

TO RETURN TO THE OPERATION MODE :

Click the Sensitivity Control Dial.

TO DOWNLOAD THE INFORMATION:

Refer to Setup Mode, 01 Data Transfer

9

Image 10
Contents R a U N C U T a BPA-9 LRM-9 CFM-9BCH-WTG2 HTS-2 Sheath LLA 5PC-MP 5PC-BNC QCH-9 Page Overview Operation ModeSET UP Mode Basic ComponentsTrisonictm Scanning Module PISTOL-GRIP Housing Display PanelPLUG-IN Modules Stethoscope Contact ModuleBattery Compartment Trigger ON/OFF Trigger SwitchConnector Diagram BatteryAccessories Sheath WTG-2SP Warble Pipe Threaded Tone GeneratorBCH-WTG2 LLASensitivity Control Dial To Adjust the SENSITIVITY/VOLUME Operation ModeBAR Graph Display Yellow Store Button To Store a Reading To Adjust the FrequencyTo Overwrite Data or to Enter Data in a NEW Location Text Editor Setup ModeTo Enter Text in the Text Field Set Time and Date DB Offset Calibration Due Date Method of Airborne Detection Users InstructionsTRISONIC- Scanning Module Stethoscope ModuleWarble Tone Generator UE-WTG-1 AUTO-SHUTDOWN Battery FeatureResetting the on Board Computer To Charge the Warble Tone GeneratorLeak Detection To Confirm a Leak HOW to Locate LeaksOvercoming Difficulties Shielding Techniques Tone Test UltratonePage Electric ARC, CORONA, Tracking Detection Monitoring Bearing Wear Detecting Bearing Failure Procedure for Bearing History HistoricalFor Comparative Test Over-Lubrication Lack of LubricationDo not OVER-LUBRICATE Slow Speed BearingsGeneral Mechanical Trouble Shooting ProcedureLocating Faulty Steam Traps General STEAM/CONDENSATE/FLASH Steam ConfirmationPage Procedure for Valve Check Abcd Method Miscellaneous Problem Areas A. Underground LeaksConfirming Valve Leakage in Noisy Pipe Systems Partial Blockage Leakage Behind WallsFlow Direction Page Instructions for Setting Combination on Carrying Case Hayes Street Elmsford, NY 10523 USA Tel