09 Exit to program
Click the Sensitivity Control dial and you will exit to Operations Mode.
6. USERS INSTRUCTIONS
TRISONIC- SCANNING MODULE:
∙Plug in to front end.
∙Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH) and plug in.
∙For general use position the frequency selection to 40 kHz.
∙Start to scan the test area.
METHOD OF AIRBORNE DETECTION:
The method of air borne detection is to go from the "gross to the fine". If there is too much ultrasound in the area, reduce the sensitivity, place the RUBBER FOCUSING PROBE (described below) over the scanning module and proceed to follow the test sound to its' loudest point constantly reducing the sensitivity and following the display
HEADSET:
To use, simply plug the headset cord into the headset Jack on the pistol housing, and place the headphones over your ears. If a hard hat is to be worn, it is recommended to use UE Systems' model
RUBBER FOCUSING PROBE :
To use, simply slip it over the front of the scanning module or the contact module.
NOTE: To prevent damage to the module plugs, always remove the module BEFORE attaching and/or removing the Rubber focusing Probe.
STETHOSCOPE MODULE:
∙Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH) and plug in.
∙Touch test area.
As with the scanning module, go from the "gross" to the "fine". Start a maximum sensitivity on the Sensitivity Selection dial and proceed to reduce the sensitivity until a satisfactory sound level is achieved. At times it may be necessary to utilize the stethoscope probe with the sensitivity level at or near maximum. Occasionally when in this situation stray ultrasound may interfere with clear reception and be confusing. If this occurs, place the RUBBER FOCUSING PROBE over the Stethoscope probe to insulate against the stray ultrasound.
STETHOSCOPE EXTENSION KIT:
1.Remove the Stethoscope Module from the Metered Pistol Housing.
2.Unscrew the metal rod in the Stethoscope Module.
3.Look at the thread of the rod you just unscrewed and locate a rod in the kit that has the same size thread - this is the "base piece".
4.Screw the Base Piece into the Stethoscope Module.
5.If all 31" (78.7 cm) are to be utilized, locate the middle piece. (This is the rod with a female fitting at one end) and screw this piece into the base piece.
6.Screw third "end piece" into middle piece.
7.If a shorter length is desired, omit step 5 and screw "end piece" into "base piece".
TO CHARGE THE UP9000:
∙Plug recharger cable into recharger jack on the UP9000 and then plug the recharger into a wall receptacle.
∙Make sure that the LED on the charger is lit when recharging.
∙The LED turns OFF when the battery is charged. The instrument may stay connected to the charger without damaging the battery.
∙WARNING: Use the supplied UE Systems recharger
14