09 Exit to program

Click the Sensitivity Control dial and you will exit to Operations Mode.

6. USERS INSTRUCTIONS

TRISONIC- SCANNING MODULE:

Plug in to front end.

Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH) and plug in.

For general use position the frequency selection to 40 kHz.

Start to scan the test area.

METHOD OF AIRBORNE DETECTION:

The method of air borne detection is to go from the "gross to the fine". If there is too much ultrasound in the area, reduce the sensitivity, place the RUBBER FOCUSING PROBE (described below) over the scanning module and proceed to follow the test sound to its' loudest point constantly reducing the sensitivity and following the display

HEADSET:

To use, simply plug the headset cord into the headset Jack on the pistol housing, and place the headphones over your ears. If a hard hat is to be worn, it is recommended to use UE Systems' model DHC-2HH hard hat headphones.

RUBBER FOCUSING PROBE :

To use, simply slip it over the front of the scanning module or the contact module.

NOTE: To prevent damage to the module plugs, always remove the module BEFORE attaching and/or removing the Rubber focusing Probe.

STETHOSCOPE MODULE:

Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH) and plug in.

Touch test area.

As with the scanning module, go from the "gross" to the "fine". Start a maximum sensitivity on the Sensitivity Selection dial and proceed to reduce the sensitivity until a satisfactory sound level is achieved. At times it may be necessary to utilize the stethoscope probe with the sensitivity level at or near maximum. Occasionally when in this situation stray ultrasound may interfere with clear reception and be confusing. If this occurs, place the RUBBER FOCUSING PROBE over the Stethoscope probe to insulate against the stray ultrasound.

STETHOSCOPE EXTENSION KIT:

1.Remove the Stethoscope Module from the Metered Pistol Housing.

2.Unscrew the metal rod in the Stethoscope Module.

3.Look at the thread of the rod you just unscrewed and locate a rod in the kit that has the same size thread - this is the "base piece".

4.Screw the Base Piece into the Stethoscope Module.

5.If all 31" (78.7 cm) are to be utilized, locate the middle piece. (This is the rod with a female fitting at one end) and screw this piece into the base piece.

6.Screw third "end piece" into middle piece.

7.If a shorter length is desired, omit step 5 and screw "end piece" into "base piece".

TO CHARGE THE UP9000:

Plug recharger cable into recharger jack on the UP9000 and then plug the recharger into a wall receptacle.

Make sure that the LED on the charger is lit when recharging.

The LED turns OFF when the battery is charged. The instrument may stay connected to the charger without damaging the battery.

WARNING: Use the supplied UE Systems recharger (BCH-9 or BCH-92) only. Use of unauthorized rechargers will void the warranty and may degrade or damage the battery.

14

Image 15
Contents R a U N C U T a LRM-9 CFM-9 BPA-9BCH-WTG2 HTS-2 Sheath LLA 5PC-MP 5PC-BNC QCH-9 Page Basic Components Operation ModeSET UP Mode OverviewStethoscope Contact Module PISTOL-GRIP Housing Display PanelPLUG-IN Modules Trisonictm Scanning ModuleBattery Trigger ON/OFF Trigger SwitchConnector Diagram Battery CompartmentAccessories LLA WTG-2SP Warble Pipe Threaded Tone GeneratorBCH-WTG2 SheathOperation Mode Sensitivity Control Dial To Adjust the SENSITIVITY/VOLUMEBAR Graph Display To Adjust the Frequency Yellow Store Button To Store a ReadingTo Overwrite Data or to Enter Data in a NEW Location Setup Mode Text EditorTo Enter Text in the Text Field Set Time and Date DB Offset Calibration Due Date Stethoscope Module Users InstructionsTRISONIC- Scanning Module Method of Airborne DetectionTo Charge the Warble Tone Generator AUTO-SHUTDOWN Battery FeatureResetting the on Board Computer Warble Tone Generator UE-WTG-1Leak Detection HOW to Locate Leaks To Confirm a LeakOvercoming Difficulties Tone Test Ultratone Shielding TechniquesPage Electric ARC, CORONA, Tracking Detection Monitoring Bearing Wear Procedure for Bearing History Historical Detecting Bearing FailureFor Comparative Test Slow Speed Bearings Lack of LubricationDo not OVER-LUBRICATE Over-LubricationProcedure General Mechanical Trouble ShootingGeneral STEAM/CONDENSATE/FLASH Steam Confirmation Locating Faulty Steam TrapsPage Procedure for Valve Check Miscellaneous Problem Areas A. Underground Leaks Abcd MethodConfirming Valve Leakage in Noisy Pipe Systems Leakage Behind Walls Partial BlockageFlow Direction Page Instructions for Setting Combination on Carrying Case Hayes Street Elmsford, NY 10523 USA Tel