Alpine 9000 instruction manual Calibration Due Date

Page 14

06 Calibration Due Date

Shown as Cal Due Date in the menu, this date is set at the factory and displays the recommended Recalibration/service date. This is one mode that cannot be changed by a user. It is only set at the factory after a service has been performed. NOTE: This cannot be changed.

07 Text Editor

Text editor will enable or disable text entry when a reading is to be saved during the operation mode. If text notes are to be manually entered, select the ON mode. If text has been preset in the Ultratrend™ software or if text entry is not needed, select OFF.

To select Text Editor:

1.Make sure the Ultraprobe is off.

2.Press (click) both the Yellow Store button and the Sensitivity dial at the same time, then squeeze and hold the trigger.

3.When in the first Menu Selection , “Data Transfer” (Menu 01), you may move to any of the other Menu Selections by spinning the Sensitivity Control up or down (clockwise or counter clockwise).

4.Spin to Text Editor Sel (Select), Menu 07 blinks.

5.Click the Sensitivity Control Dial to enter the Text Editor enable mode.

6.Spin the Sensitivity Control Dial to select OFF or ON.

7.Click the Sensitivity Control Dial to set and return to Set Up Mode

8.Spin to Exit to PGM (Exit to Program) Menu 09 blinks. Click to enter Operation Mode

Menu

07

Select

Text

Editor

 

 

 

 

click

Text Editor Sel. Text Editor = (OFF)

spin

Text Editor Sel. Text Editor = (ON)

08 Factory Defaults

This mode allows users to retain or to delete the information stored in the instrument and to restore the factory default settings of the instrument. Confirm YES means that the onboard computer will default to original factory settings and all stored data will be deleted. Confirm NO retains all stored data and current instrument settings.

The Factory Default Settings: ∙Maximum Sensitivity ∙Frequency = 40 kHz ∙Display Mode = Real Time ∙dB Scale = Relative ∙Offset Value = 0

∙Text Editor = ON

∙Peak Value Indicator ( bar graph ) = 0

To select Factory Defaults:

1.Make sure the Ultraprobe is off.

2.Press (click) both the Yellow Store button and the Sensitivity dial at the same time, then squeeze and hold the trigger.

3.When in the first Menu Selection : “Data Transfer” (Menu 01), you may move to any of the other Menu Selections by spinning the Sensitivity Control up or down (clockwise or counter clockwise).

4.Spin to Factory Defaults ( Menu 08 Blinks).

5.Click to enter the Factory Default Set Up Mode

6.Spin up or down to select either YES or NO.

7.Click to set and return to Set Up Mode.

8.Spin to Exit to PGM (Exit to Program) Menu 09 blinks. Click to enter Operation Mode

Menu

08

Select

Factory

Defaults

 

click

Factory Defaults Confirm = (NO)

spin

Factory Defaults Confirm = (YES)

13

Image 14
Contents R a U N C U T a BCH-WTG2 HTS-2 Sheath LLA 5PC-MP 5PC-BNC QCH-9 LRM-9 CFM-9BPA-9 Page Overview Operation ModeSET UP Mode Basic ComponentsTrisonictm Scanning Module PISTOL-GRIP Housing Display PanelPLUG-IN Modules Stethoscope Contact ModuleBattery Compartment Trigger ON/OFF Trigger SwitchConnector Diagram BatteryAccessories Sheath WTG-2SP Warble Pipe Threaded Tone GeneratorBCH-WTG2 LLABAR Graph Display Operation ModeSensitivity Control Dial To Adjust the SENSITIVITY/VOLUME To Overwrite Data or to Enter Data in a NEW Location To Adjust the FrequencyYellow Store Button To Store a Reading To Enter Text in the Text Field Setup ModeText Editor Set Time and Date DB Offset Calibration Due Date Method of Airborne Detection Users InstructionsTRISONIC- Scanning Module Stethoscope ModuleWarble Tone Generator UE-WTG-1 AUTO-SHUTDOWN Battery FeatureResetting the on Board Computer To Charge the Warble Tone GeneratorLeak Detection Overcoming Difficulties HOW to Locate LeaksTo Confirm a Leak Shielding Techniques Tone Test UltratonePage Electric ARC, CORONA, Tracking Detection Monitoring Bearing Wear For Comparative Test Procedure for Bearing History HistoricalDetecting Bearing Failure Over-Lubrication Lack of LubricationDo not OVER-LUBRICATE Slow Speed BearingsGeneral Mechanical Trouble Shooting ProcedureLocating Faulty Steam Traps General STEAM/CONDENSATE/FLASH Steam ConfirmationPage Procedure for Valve Check Confirming Valve Leakage in Noisy Pipe Systems Miscellaneous Problem Areas A. Underground LeaksAbcd Method Flow Direction Leakage Behind WallsPartial Blockage Page Instructions for Setting Combination on Carrying Case Hayes Street Elmsford, NY 10523 USA Tel