Section 3 | Installation |
DANGER
Only qualified personnel should conduct the tasks described in this section of the manual.
DANGER
Seul un technicien qualifié peut effectuer les tâches d'installation décrites dans cette section du manuel.
The Combination pH/ORP Sensor can be used with either an sc100 or sc1000 controller. Refer to section 3.2 on page 13 for sc100 installation instructions and section 3.3 on page 16 for sc1000 installation instructions.
The sc sensor should be wired to the digital gateway before connecting the digital gateway to the sc100 or sc1000 Controller. The digital gateway is designed to provide a digital interface to the appropriate digital controller. Refer to section 3.1 for more information.
3.1 Wiring the sc Sensor to the Digital Gateway
DANGER
The sc100 and certain versions of the sensor are suitable for use in Class 1, Division 2, Groups A, B, C, D Hazardous Locations . See Control Drawing
DANGER
Le sc100 et certaines versions du capteur peuvent être utilisés dans des endroits dangereux de la Classe 1, Division 2, Groupes A, B, C, D.
1.Route the cable from the sensor through the strain relief in the digital gateway then properly terminate the wire ends (see Figure 5).
Note: Do not tighten the strain relief until the digital gateway is wired and the two halves are threaded securely together.
2.Insert the wires as shown in Table 3 and Figure 6.
3.Make sure the
4.Tighten the strain relief to secure the sensor cable.
5.Connect the digital gateway to the controller.
•sc100
•sc100 Hazardous
•sc1000 Connection