Operation
5.5ORP Sensor Setup Menu
SELECT SENSOR (if more than one sensor is attached)
CALIBRATE (Main Menu Item)
| 1 POINT MANUAL | Calibration against a single known sample. |
|
|
|
| TEMP ADJUST | Adjust the displayed temperature by up to ± 15 °C. |
|
|
|
| DEFAULT SETUP | Restores the system to the original factory calibration. |
|
|
|
CONFIGURE |
| |
|
|
|
| EDIT NAME | Enter up to a |
|
|
|
| SELECT SENSOR | Choose from the displayed sensor type (pH or ORP). |
|
|
|
| TEMP UNITS | Choose from the displayed options (°C or °F). |
|
|
|
| LOG SETUP | Choose SENSOR INTERVAL to set the sensor log interval or select TEMP INTERVAL to set the |
| temperature log interval. | |
|
| |
|
|
|
| AC FREQUENCY | Choose 50 or 60 Hz depending on the power line frequency for optimal noise rejection. Default is |
| 60 Hz. | |
|
| |
|
|
|
| FILTER | Select |
|
|
|
| TEMP ELEMENT | Select type of temperature element from the displayed choices. |
|
|
|
| CAL DAYS | Number of days since the last calibration. Default notification at 60 days. |
|
|
|
| SENSOR DAYS | Number of days the sensor has been in operation. Default notification at 365 days. |
|
|
|
| IMPED LIMITS | Set min/max electrode sensor impedance limits. |
|
|
|
| DEFAULT SETUP | Resets all |
|
|
|
DIAG/TEST |
| |
|
| Display the sensor type, entered name of the sensor (Default: Digital Gateway serial number and |
| PROBE INFO | name), the sensor serial number, the software version number, and the sensor driver version |
|
| number. |
|
|
|
| CAL DATA | Displays the slope and the date of the last calibration. |
|
|
|
|
| SENSOR SIGNAL: displays the sensor output in mV |
|
| SENSOR ADC COUNTS: displays the sensor ADC counts |
|
| TEMP ADC COUNTS: shows raw data for temperature ADC counts. ADC counts are comparable |
|
| to A/D counts and are for sensor electronic diagnostic purposes only. |
| SIGNALS | ELECTRODE STATE: Identifies the state of the electrode (good or bad) depending on whether |
|
| the impedance is within preset limits. |
|
| ACTIVE ELECT: Shows the impedance (Mohms) of the active electrode if Imped Status is set to |
|
| Enabled. |
|
| IMPED STATUS: Sensor diagnostic. Choose Enabled or Disabled. |
|
|
|
|
| SENSOR DAYS: displays the cumulative days the sensor has been in use. |
| COUNTERS | RESET SENSOR: allows the sensor counter to be reset to zero. |
|
| ELECTRODE DAYS: Cumulative days the electrode has been in use. |
|
|
|