Operation

5.5ORP Sensor Setup Menu

SELECT SENSOR (if more than one sensor is attached)

CALIBRATE (Main Menu Item)

 

1 POINT MANUAL

Calibration against a single known sample.

 

 

 

 

TEMP ADJUST

Adjust the displayed temperature by up to ± 15 °C.

 

 

 

 

DEFAULT SETUP

Restores the system to the original factory calibration.

 

 

 

CONFIGURE

 

 

 

 

 

EDIT NAME

Enter up to a 10-digit name in any combination of symbols and alpha or numeric characters.

 

 

 

 

SELECT SENSOR

Choose from the displayed sensor type (pH or ORP).

 

 

 

 

TEMP UNITS

Choose from the displayed options (°C or °F).

 

 

 

 

LOG SETUP

Choose SENSOR INTERVAL to set the sensor log interval or select TEMP INTERVAL to set the

 

temperature log interval.

 

 

 

 

 

 

AC FREQUENCY

Choose 50 or 60 Hz depending on the power line frequency for optimal noise rejection. Default is

 

60 Hz.

 

 

 

 

 

 

FILTER

Select 0–60 second signal averaging time.

 

 

 

 

TEMP ELEMENT

Select type of temperature element from the displayed choices.

 

 

 

 

CAL DAYS

Number of days since the last calibration. Default notification at 60 days.

 

 

 

 

SENSOR DAYS

Number of days the sensor has been in operation. Default notification at 365 days.

 

 

 

 

IMPED LIMITS

Set min/max electrode sensor impedance limits.

 

 

 

 

DEFAULT SETUP

Resets all user-editable options to their factory-defaults.

 

 

 

DIAG/TEST

 

 

 

Display the sensor type, entered name of the sensor (Default: Digital Gateway serial number and

 

PROBE INFO

name), the sensor serial number, the software version number, and the sensor driver version

 

 

number.

 

 

 

 

CAL DATA

Displays the slope and the date of the last calibration.

 

 

 

 

 

SENSOR SIGNAL: displays the sensor output in mV

 

 

SENSOR ADC COUNTS: displays the sensor ADC counts

 

 

TEMP ADC COUNTS: shows raw data for temperature ADC counts. ADC counts are comparable

 

 

to A/D counts and are for sensor electronic diagnostic purposes only.

 

SIGNALS

ELECTRODE STATE: Identifies the state of the electrode (good or bad) depending on whether

 

 

the impedance is within preset limits.

 

 

ACTIVE ELECT: Shows the impedance (Mohms) of the active electrode if Imped Status is set to

 

 

Enabled.

 

 

IMPED STATUS: Sensor diagnostic. Choose Enabled or Disabled.

 

 

 

 

 

SENSOR DAYS: displays the cumulative days the sensor has been in use.

 

COUNTERS

RESET SENSOR: allows the sensor counter to be reset to zero.

 

 

ELECTRODE DAYS: Cumulative days the electrode has been in use.

 

 

 