Installation
Figure 7 Attaching the Digital Gateway using the Quick-connect Fittings
3.2.1.2 Hard-wiring the Digital Gateway to the sc100
Important Note:
1.Disconnect power to the controller if powered.
2.Open the controller cover.
3.Disconnect and remove the existing wires between the
4.Remove the
5.Cut the connector from the sensor cable.
6.Strip the insulation on the cable back
Note: Use of strain relief fitting other than Cat. No. 16664 may result in a hazard. Use only the recommended strain relief fitting.
7.Pass the cable through conduit and a conduit hub or a strain relief fitting (Cat. No. 16664) and an available access hole in the controller enclosure. Tighten the fitting.
8.Reinstall the plug on the sensor access opening to maintain the environmental rating.
9.Wire as shown in Table 4 and Figure 8.
10.Close and secure the cover.
14