General Information

2.2 General Sensor Information

Optional equipment, such as mounting hardware for the probe, is supplied with instructions for all user installation tasks. Several mounting options are available, allowing the probe to be adapted for use in many different applications.

2.2.1 Sensor Body Styles

Combination pH and ORP sensors are available in three body styles:

Convertible Body Style (Figure 1 and Figure 2)— has ¾-in. NPT threads at both ends of the body for mounting in any of the following configurations:

into a standard ¾-in. NPT pipe tee

onto the end of a pipe for immersion into a vessel

Insertion Body Style (Figure 3)— for mounting into the pipe adapter of a ball valve hardware assembly. This hardware enables the sensor to be inserted into or retracted from the process without stopping the process flow.

Sanitary Body Style (Figure 4)— features a built-in 2-in. flange for mounting into a 2-in. sanitary tee. Included with the sanitary-style sensor is a special cap and EDPM compound gasket for use with the sanitary hardware.

Figure 1 Convertible Style Sensor with Flat Electrode

¾-inch NPT

Typical

22.2 mm (0.875 inch)

22.9mm (0.90 inch) 149.9 mm (5.9 inches)

7.62mm

(1.30 inches)

Figure 2 Convertible-style Sensor with Dome Electrode

¾-inch NPT

Typical

22.2 mm (0.875 inch)

22.9mm (0.90 inch) 149.9 mm (5.9 inches)

7.62mm

(1.30 inches)

8

Page 8
Image 8
Hach 6120118 user manual General Sensor Information, General Information, Sensor Body Styles