Section 1 Specifications
Specifications are subject to change without notice.
Table 1 Combination pH and ORP Sensor Specifications
Components | ||
|
| |
Measuring Range (pH) | 0 to 14 pH | |
|
| |
Measuring Range (ORP) | ||
|
| |
Measuring Range (Temperature) | 0 to 105 °C (32 to 221 °F) | |
|
| |
Probe Operating Temperature | 0 to 105 °C (32 to 221 °F) | |
|
| |
Probe Storage Temperature | ||
|
| |
Temperature Compensation | pH: Pt 1000 ohm RTD | |
ORP: N/A | ||
| ||
|
| |
Accuracy (Analyzer only) | pH: 0.1% of span | |
ORP: Limited to calibration solution accuracy (± 20 mV) | ||
| ||
|
| |
Stability (Analyzer only) | 0.05% or span per 24 hours, | |
|
| |
Repeatability (Analyzer only) | 0.1% of span or better | |
|
| |
Temperature Accuracy (Analyzer only) | ±0.5 °C (±0.9 °F) | |
|
| |
Temperature Drift (Analyzer only) | Zero and Span: less than 0.03% of span per °C | |
|
| |
Calibration Methods (Analyzer only) | Two point automatic, one point automatic, two point manual, one point manual | |
|
| |
Maximum Probe Immersion Depth/ | Submersible to 107 m (350 ft)/1050 kPa (150 psi) | |
Pressure | ||
| ||
|
| |
Sensor Interface | Modbus from digital gateway | |
|
| |
Probe Cable Length | 4.6 m (15 ft) | |
|
| |
Probe Weight | Depends on selected sensor | |
|
| |
Probe Dimensions | See Figure 1 on page 8 through Figure 4 on page 9. | |
|
|
| Table 2 Digital Gateway Specifications | |
|
|
|
Weight |
| 145 g (5 oz) |
|
|
|
Dimensions |
| 17.5 x 3.4 cm (7 x 13/8 in.) |
Operating Temperature |
| |
|
|
|
5