Alpine 9000 instruction manual Set Time and Date

Page 12

02Set Time and Date

1.Make sure the Ultraprobe is off.

2.Press (click) both the Yellow Store button and the Sensitivity dial at the same time, then squeeze and hold the trigger.

3.When in the first Menu Selection : “Data Transfer” (Menu 01), you may move to any of the other Menu Selections by spinning the Sensitivity Control up or down (clockwise or counter clockwise).

4.Spin to “Set Time and Date” (Menu 02 blinks) and click in (EXIT Blinks), .

5.Spin to desired month or day or year and Click (selected number will blink rapidly).

6.Spin to select a new value

7.Click to set.

8.Spin to TIME setting and click on either Hour or Minute (the displayed number will blink rapidly).

9.Once an hour or minute has been selected, spin to select a new value.

10.Click to set.

11.When through, Spin the Sensitivity Control until EXIT flashes.

12.Click the Sensitivity Control again and return to the Set Up Mode.

13.Spin to Exit to PGM (Exit to Program) Menu 9 blinks. Click to enter Operation Mode.

Menu 02 Select

Set Time & Date

click

TIME 15:30 EXIT DATE 1/01/99

03 dB Scale Select

dB Select has two settings from which to choose. These settings will determine the baseline dB reference level of the instrument. Once selected, all test results will be based on the selected baseline dB level. There are two scales: Relative and dB offset.

Relative sets the instrument to the 0 dB of the instrument’s internal minimal detection value and is the factory default setting.

dB offset is a dB level that is a new minimum reference level set by the user. This value may be any dB level above the natural 0 dB of the instrument. Once set, the preset level must be subtracted from the reading to determine an accurate dB increase. (EG: if “10” is the dB offset value and a subsequent reading is 25 dB, then the increase is 15 dB.)

To select a dB reference scale:

1.Make sure the Ultraprobe is off.

2.Press (click) both the Yellow Store button and the Sensitivity dial at the same time, then squeeze and hold the trigger.

3.When in the first Menu Selection : “Data Transfer” (Menu 01), you may move to any of the other Menu Selections by spinning the Sensitivity Control up or down (clockwise or counter clockwise).

4.Spin to dB Scale Select (Menu 03 blinks).

6.Click in the Sensitivity Control.

7.Spin the Sensitivity Control to the desired scale (Relative or Offset).

8.Click the Sensitivity Control to set and return to the set up mode.

9.Spin to Exit to PGM (Exit to Program) Menu 09 blinks. Click to enter Operation Mode

Menu

03

Select

dB

Scale

Select

click

dB Scale Select Relative

spin

dB Scale Select Offset

11

Image 12
Contents R a U N C U T a LRM-9 CFM-9 BPA-9BCH-WTG2 HTS-2 Sheath LLA 5PC-MP 5PC-BNC QCH-9 Page Operation Mode SET UP ModeOverview Basic ComponentsPISTOL-GRIP Housing Display Panel PLUG-IN ModulesTrisonictm Scanning Module Stethoscope Contact ModuleTrigger ON/OFF Trigger Switch Connector DiagramBattery Compartment BatteryAccessories WTG-2SP Warble Pipe Threaded Tone Generator BCH-WTG2Sheath LLAOperation Mode Sensitivity Control Dial To Adjust the SENSITIVITY/VOLUMEBAR Graph Display To Adjust the Frequency Yellow Store Button To Store a ReadingTo Overwrite Data or to Enter Data in a NEW Location Setup Mode Text EditorTo Enter Text in the Text Field Set Time and Date DB Offset Calibration Due Date Users Instructions TRISONIC- Scanning ModuleMethod of Airborne Detection Stethoscope ModuleAUTO-SHUTDOWN Battery Feature Resetting the on Board ComputerWarble Tone Generator UE-WTG-1 To Charge the Warble Tone GeneratorLeak Detection HOW to Locate Leaks To Confirm a LeakOvercoming Difficulties Shielding Techniques Tone Test UltratonePage Electric ARC, CORONA, Tracking Detection Monitoring Bearing Wear Procedure for Bearing History Historical Detecting Bearing FailureFor Comparative Test Lack of Lubrication Do not OVER-LUBRICATEOver-Lubrication Slow Speed BearingsGeneral Mechanical Trouble Shooting ProcedureLocating Faulty Steam Traps General STEAM/CONDENSATE/FLASH Steam ConfirmationPage Procedure for Valve Check Miscellaneous Problem Areas A. Underground Leaks Abcd MethodConfirming Valve Leakage in Noisy Pipe Systems Leakage Behind Walls Partial BlockageFlow Direction Page Instructions for Setting Combination on Carrying Case Hayes Street Elmsford, NY 10523 USA Tel