4. OPERATION MODE

DISPLAY PANEL:

When the trigger is pressed to turn the instrument on, the Display Panel will display intensity levels simultaneously on a bar graph and as a numerical decibel value. The current selected frequency will also be shown. Remaining Battery Charge is shown in the upper right corner. The letters R, S or P will alternate with the battery indicator in the upper right corner. R indicates that the instrument is running in “Real Time”, S indicates “Snap Shot” and P indicates “Peak Hold”. Should the instrument be set in the Offset Mode, then the letters RO, SO and PO will be displayed.

BAR GRAPH DISPLAY:

The bar graph has 16 segments. Each segment represents 3 decibels. At the end of the bar graph is a vertical line, which indicates the peak intensity. This is a Peak Hold function. When in operation, the bar graph will move up and down the scale as an indication of the amplitude of a sensed ultrasound. The Peak Hold indicator will remain at the highest sensed intensity during a particular inspection until: 1. A new maximum reading is detected, or 2. The trigger is released and the instrument is turned off. At which time it will reset.

06 dB 40 kHz R

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _

06 dB 40 kHz S

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

06 dB

_ _ _ _ _

40 kHz P

_ _ _ _ _ _ _ _ _ _

Real Time = R Flashes

Snap Shot = S Flashes

Peak Hold = P Flashes

 

All of these alternate with battery level

Indicator

 

 

 

SENSITIVITY CONTROL DIAL:

TO ADJUST THE SENSITIVITY/VOLUME:

Look at the meter. If the instrument is within range, the dB decibel indicator must blink. The kHz ( frequency ), indicator must be steady and not blink.

If the frequency indicator is blinking, click in the sensitivity control dial until the frequency indicator is steady and the decibel indicator blinks.. This indicates that you are now able to adjust the sensitivity.

Once in the Sensitivity mode, turn the Sensitivity control dial clockwise to increase the sensitivity and counter clockwise to decrease the sensitivity.

The Sensitivity control dial increases/decreases the sensitivity of the instrument simultaneously with the sound level in the headphones

NOTE: the instrument needs to be in range for accurate testing.

If the sensitivity is too low, a blinking arrow pointing to the right will appear and there will be no numeric decibel visible in the display panel. If this occurs, increase the sensitivity until the arrow disappears (in low level sound environments the arrow will blink continuously and It will not be possible to achieve a dB indication until a higher intensity level is sensed).

If the sensitivity is too high, a blinking arrow pointing to the left will appear and there will be no numeric decibel visible on the display panel. Reduce the sensitivity until the arrow disappears and the numeric decibel value is shown.

NOTE: The blinking arrow indicates the direction in which the Sensitivity Control Dial is to be turned.

The Sensitivity Control Dial controls the bar graph display.

Each click of the sensitivity dial changes the sensitivity / volume by 1 dB

8

Image 9
Contents R a U N C U T a LRM-9 CFM-9 BPA-9BCH-WTG2 HTS-2 Sheath LLA 5PC-MP 5PC-BNC QCH-9 Page SET UP Mode Operation ModeOverview Basic ComponentsPLUG-IN Modules PISTOL-GRIP Housing Display PanelTrisonictm Scanning Module Stethoscope Contact ModuleConnector Diagram Trigger ON/OFF Trigger SwitchBattery Compartment BatteryAccessories BCH-WTG2 WTG-2SP Warble Pipe Threaded Tone GeneratorSheath LLAOperation Mode Sensitivity Control Dial To Adjust the SENSITIVITY/VOLUMEBAR Graph Display To Adjust the Frequency Yellow Store Button To Store a ReadingTo Overwrite Data or to Enter Data in a NEW Location Setup Mode Text EditorTo Enter Text in the Text Field Set Time and Date DB Offset Calibration Due Date TRISONIC- Scanning Module Users InstructionsMethod of Airborne Detection Stethoscope ModuleResetting the on Board Computer AUTO-SHUTDOWN Battery FeatureWarble Tone Generator UE-WTG-1 To Charge the Warble Tone GeneratorLeak Detection HOW to Locate Leaks To Confirm a LeakOvercoming Difficulties Tone Test Ultratone Shielding TechniquesPage Electric ARC, CORONA, Tracking Detection Monitoring Bearing Wear Procedure for Bearing History Historical Detecting Bearing FailureFor Comparative Test Do not OVER-LUBRICATE Lack of LubricationOver-Lubrication Slow Speed BearingsProcedure General Mechanical Trouble ShootingGeneral STEAM/CONDENSATE/FLASH Steam Confirmation Locating Faulty Steam TrapsPage Procedure for Valve Check Miscellaneous Problem Areas A. Underground Leaks Abcd MethodConfirming Valve Leakage in Noisy Pipe Systems Leakage Behind Walls Partial BlockageFlow Direction Page Instructions for Setting Combination on Carrying Case Hayes Street Elmsford, NY 10523 USA Tel