Fluke PN 643657 instruction sheet Specifications, Function Range Resolution

Page 13

Specifications

Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with relative humidity to 90%. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave input. Accuracy specifications are given as follows:

±([% of reading] + [number of least significant digits])

Function

Range

Resolution

Temperature

-10°C to

400°C

0.1°C or

14°F to

752°F

0.2°F

 

 

 

 

(Type K Thermocouple)

-40°C to -10°C

0.1°C or

 

-40°F to

14°F

0.2°F

 

 

 

 

Error does not include Type K Thermocouple errors.

Accuracy

±(1.0% + 0.8°C)

typical

±(1.0% + 1.5°F) typical

±(5.0% + 1.5°C)

typical

±(5.0% + 3.3°F) typical

Functionf

(50 to 400 Hz)

E

Ω

ML

Range

4000 mV1

4.000V

40.00V

400.0V

600V

4000 mV1

4.000V

40.00V

400.0V

600V

400.0Ω

4.000 kΩ

40.00kΩ

400.0kΩ

4.000 MΩ

40.00MΩ

1.000 μF

10.00μF

100.0μF 10000 μF

Resolution

1mV

0.001V

00.01V

000.1V

1V

1mV

0.001V

00.01V

000.1V

1V

0.1Ω

0.001 kΩ

0.01 kΩ

0.1 kΩ

0.001 MΩ

0.01 MΩ

0.001 μF

0.01 μF

0.1μF

1 μF

Accuracy

±(1.9% + 3)

±(1.9% + 3)

±(1.9% + 3)

±(1.9% + 3)

±(1.9% + 3)

±(0.9% + 2)

±(0.9% + 2)

±(0.9% + 1)

±(0.9% + 1)

±(0.9% + 1)

±(0.9% + 2)

±(0.9% + 1)

±(0.9% + 1)

±(0.9% + 1)

±(0.9% + 1)

±(1.5% + 3)

±(1.9% + 2)

±(1.9% + 2)

±(1.9% + 2)

1000 μF ±(1.9% + 2)

>1000 μF ±(10% + 90) typical

2.000V

0.001V

±(1.9% + 2)2

1.The 4000 mV range can be entered only in manual range mode. Use the 4000 mV range with accessories.

2.The beeper is guaranteed to come on at <25Ω and turn off at >250Ω. The meter detects opens or shorts 250 μs.

Image 13
Contents Read First Safety Information Multimeter with TemperatureBattery Saver SymbolsAutomatic Range Selection Manual Range SelectionResistance and Continuity eR AC and DC Voltage∙Check Disable and re-enable VCheckTemperature If the capacitor requires more Capacitance EMicroamps μA Typical voltage Diode Test GRecords the lowest and highest measurements MIN MAX μFMIN MAX with Elapsed Time μF Continuity Capture Disabling the BeeperReplacement Parts MaintenanceFunction Range Resolution SpecificationsMIN MAX Recording Accuracy and Response Time MIN MAX Recording with Elapsed Time Limited Warranty & Limitation of Liability