Fluke PN 643657 instruction sheet MIN MAX Recording Accuracy and Response Time

Page 14

Function

(50 Hz to 400

Hz)

Range

Resolution

Accuracy

0 to 200 μA

0.1 μA

±(2% + 3 counts)

 

 

 

0 to 200 μA

0.1 μA

±(1% + 2 counts)

Burden Voltage

<5 mV/μA

<5 mV/μA

Functionf

E

Overload

Protection1

600V rms

600V rms

Input Impedance (Nominal)

>5 MΩ <100 pF

VCheck and LoZ = >2 kΩ <200 pF (ac coupled)2

>10MΩ <100 pF

VCheck and LoZ = >2 kΩ <200 pF2

f

E600V rms

600V rms

600V rms

600V rms

Common Mode Rejection Ratio (1 kΩ Unbalanced)

>60 dB at dc 50 or 60 Hz

>100 dB at dc, 50 or 60 Hz

Open Circuit Test Voltage

<1.5V dc

2.4-3.0V dc

Normal Mode Rejection

>50 dB at 50 Hz or 60 Hz

Full Scale Voltage

To 4.0 MΩ

40 MΩ

<450 mV dc

<1.5V dc

2.400V dc

J

L600V rms

600V rms

1. 3 x 106 V Hz maximum

Short Circuit Current

<500 μA

0.95 mA (typical)

2.2 kΩ input impedance up to 50V. Impedance increases with input voltage to >300 kΩ at 600V.

MIN MAX Recording Accuracy and Response Time

Specified accuracy of the measurement function ±12 digits in dc for changes >200 ms in duration (±40 digits in ac). Typical 100 ms response to 80%.

Example 1: This would mean ±1.2° when recording temperature.

Example 2: This would mean ±12 μA when recording μA or ±12A if used with a dc amp probe (with a mV input).

Image 14
Contents Multimeter with Temperature Read First Safety InformationAutomatic Range Selection SymbolsBattery Saver Manual Range SelectionAC and DC Voltage Resistance and Continuity eRDisable and re-enable VCheck ∙CheckTemperature Capacitance E If the capacitor requires moreMicroamps μA Diode Test G Typical voltageMIN MAX μF Records the lowest and highest measurementsMIN MAX with Elapsed Time μF Disabling the Beeper Continuity CaptureMaintenance Replacement PartsSpecifications Function Range ResolutionMIN MAX Recording Accuracy and Response Time MIN MAX Recording with Elapsed Time Limited Warranty & Limitation of Liability