Chapter 1 - Installation and Configuration

Software Configuration

VEE Runtime I/O Configuration

You must complete these steps for the Sofware to communicate with your network analyzer!.

1. Click Start -> Programs->Agilent VEE pro 6.0 Runtime-> I/O Config

2. Turn on you Network Analyzer and Click the “Find Instruments button.

3. You should see your network analyzer appear in the list

The instrument number consists of the GPIB controller ID number and the Instrument number

Example: ID (newinstrument@1416) 14=Nat inst

GPIB controller 16=Analyzer ID #

4.Select your analyzer from the list and Click on the “Properties button” You should see the window to the right, Change the name that is shown to one from the list below that matches your analyzer.

Network Analyzer

“Name” setting

HP 8510 series Analyzers

HP8510

HP 8714 series Analyzers

HP8714

HP 8720 series Analyzers

HP8720

HP 8753 series Analyzers

HP8753

HP 5071 series Analyzers

HP5071

Anritsu Scorpion Analyzers

Scorpion

Rhode & Schwarz RS series

RSZVR

Signal Generator

“Name” setting

HP 83650B Signal Gen.

HP836

HP 8350 Sweep Generator

HP8350

Spectrum Analyzer

“Name” setting

HP 8565 Series

HP8565

Power Meter

HP 436A Power Meter

HP 437B Power Meter

Anritsu ML2438A

Anritsu ML4803A

ELVA-1 DPM 10

“Name” setting HP436

HP437

ML2438

ML4803

DPM10

5.Press the “Advanced” button to bring up the Advanced Properties Dialog Box.

Set the Instrument Timeout to 15 seconds. Press OK , then OK, then “SAVE”. You have now successfully completed the Analyzer Configuration Process.

18

Image 18
Contents Desktop Antenna Measurement System Model Page Table Of Contents Page Page Introduction Features Key FeaturesRequirements Minimum System RequirementsUSB Port Platform Controller Unit Platform Controller rear view Platform Controller top viewParallel Port Platform Controller Unit Control Unit Front View Control Unit Side ViewPage Package Contents Unpacking the SystemTripod Assembly Tripod and Elevation Unit AssemblyVertical Assembly Platform Assembly Connecting the platform and vertical actuatorCable Connections Attaching the cablesSoftware and Hardware Installation Installing the SoftwareOverview Software ConfigurationLPT Parallel Port Driver Installation Instructions StepSpectrum Analyzer Name setting Network Analyzer Name settingSignal Generator Name setting Power MeterTripod Compatibility Information System Test Platform Positioning TestScreen Sleepers and Power Saving options Installation NotesSpecial Installation Notes Other Running ApplicationsPage Chapter Source Instrument Settings Only instrument Measurement SettingsMeasurement Settings Instrument Selection and Settings Receive Instrument -Power Meter SettingsMeasurement Settings Elevation setup Measurement SettingsAzimuth setup Platform Settings and CalibrationCalibration Settings Calibration Settings Scalar CalibrationPower Sensor Calibration Measurement Controls Measurement ControlsMeasurement Status and Displays Meaurement Status and Displays Measurement Status windowGraphical Status Displays Data Processing Data Processing IntroductionData Registers Active RegisterData Processing Dynamic Amplitude Plot Measurement Calculator Dipole Link SimulatorVertical Swing Correction Page System Calibration Performing a Scalar CalibrationBasic Measurements Making a basic Horizontal/Azimuth MeasurementMaking a basic Vertical/Elevation Measurement Az/El Scan Measurements Performing an AZ/EL Scan MeasurementPage Data Processing Screenshot IntroductionData Processing Feature To Label a Plot Viewing and working with measurements in 3DTo Print a Plot To Export as BMP Image File- Click File ExportViewing and working with the Dynamic Amplitude Plot InstructionsDynamic Amplitude Plot Screenshot Viewing and working with the Polar Plot display Polar Plot ScreenshotUnderstanding the Register Math Functions Using the Measurement Calculator / Register Math FunctionRegister Math / Calculator Screenshot Math FunctionsLinear LinMag LogMagData Processing Feature Exporting DataMeasurement Export Process Measurement Export Process Contd Using the Dipole and Isotropic Link Calculator Using the Dipole and Isotropic Link Calculator Contd Group Delay Screenshot Page Page Troubleshooting and Service Troubleshooting OperationTroubleshooting Service and Contact Info