HP 6000 manual System Test, Platform Positioning Test

Page 20

Chapter 1 - Installation and Configuration

System Test

Calibrating/Setting the Vertical Movement to 0 Degrees -- See Also Vertical Calibration Settings

In order for the software to move the platform correctly on the vertical axis you must have the platform at 0 degrees when starting the DAMS Softtware - or the platform must match the vertical position shown in the software before moving vertically.

If the Platform Angle and Software Angle do not match, Use the Jog buttons to move the platform to the level position, you can change the Change jog distance by using the pulldown.

Platform Positioning Test.

1.Turn on the Platform Controller

2.Start the rotator software, Click Start -> Programs -> DAMS ->DAMS x.xx

3.Click the orange “Manual Move” button located in the upper left hand section of the software. The platform should now horzontally move the number of times specified in the “Total Number of Measurements” window completing a 360 degree sweep.

4.Set the Vertical Differential slider to a value of -5 degrees and press the green “Manual Move” button, the platform should now move to -5 degrees. Slide the slider back to 0 and press the “Manual Move” button. the platform should now return to it’s starting position.

****If the Rotator Unit successfully performed the above functions, the rotator platform is fully functional and is ready for Use.****

6.If The Rotator Unit does not Perform one or more of the above tests please refer to the Troubleshooting Section located on Page 43

20

Image 20
Contents Desktop Antenna Measurement System Model Page Table Of Contents Page Page Introduction Features Key FeaturesRequirements Minimum System RequirementsUSB Port Platform Controller Unit Platform Controller rear view Platform Controller top viewParallel Port Platform Controller Unit Control Unit Front View Control Unit Side ViewPage Package Contents Unpacking the SystemVertical Assembly Tripod AssemblyTripod and Elevation Unit Assembly Platform Assembly Connecting the platform and vertical actuatorCable Connections Attaching the cablesSoftware and Hardware Installation Installing the SoftwareSoftware Configuration LPT Parallel Port Driver Installation InstructionsOverview StepNetwork Analyzer Name setting Signal Generator Name settingSpectrum Analyzer Name setting Power MeterTripod Compatibility Information System Test Platform Positioning TestInstallation Notes Special Installation NotesScreen Sleepers and Power Saving options Other Running ApplicationsPage Chapter Measurement Settings Measurement Settings Instrument Selection and SettingsSource Instrument Settings Only instrument Receive Instrument -Power Meter SettingsMeasurement Settings Measurement Settings Azimuth setupElevation setup Platform Settings and CalibrationPower Sensor Calibration Calibration SettingsCalibration Settings Scalar Calibration Measurement Controls Measurement ControlsGraphical Status Displays Measurement Status and DisplaysMeaurement Status and Displays Measurement Status window Data Processing Data Processing IntroductionData Registers Active RegisterData Processing Dynamic Amplitude Plot Vertical Swing Correction Measurement CalculatorDipole Link Simulator Page System Calibration Performing a Scalar CalibrationMaking a basic Vertical/Elevation Measurement Basic MeasurementsMaking a basic Horizontal/Azimuth Measurement Az/El Scan Measurements Performing an AZ/EL Scan MeasurementPage Data Processing Screenshot IntroductionData Processing Feature Viewing and working with measurements in 3D To Print a PlotTo Label a Plot To Export as BMP Image File- Click File ExportDynamic Amplitude Plot Screenshot Viewing and working with the Dynamic Amplitude PlotInstructions Viewing and working with the Polar Plot display Polar Plot ScreenshotUsing the Measurement Calculator / Register Math Function Register Math / Calculator ScreenshotUnderstanding the Register Math Functions Math FunctionsLinear LinMag LogMagMeasurement Export Process Data Processing FeatureExporting Data Measurement Export Process Contd Using the Dipole and Isotropic Link Calculator Using the Dipole and Isotropic Link Calculator Contd Group Delay Screenshot Page Page Troubleshooting Troubleshooting and ServiceTroubleshooting Operation Service and Contact Info