Bacharach INSIGHT manual Fyrite Insight Features & Benefits

Page 4

Fyrite® INSIGHT

1.2 Fyrite® INSIGHT Features & Benefits

AUTOMATIC ZERO - The instrument will automatically zero all sensing channels on ambient air when it is powered on. If a particular sensor is in error during warm-up, the instrument will automatically display the error and continue to operate with the sensor in error. However, all information dependent on the sensor in error will not be presented.

TEMPERATURE UNITS - Temperatures can be displayed in Centigrade or Fahrenheit.

PRESSURE UNITS - Pressures can be displayed in milliBar (mB), Pascal (Pa), hectoPascal (hPa), Millimeters of Water Column (mmwc), or Inches of Water Column (inwc).

LANGUAGE - The display options are English, French or Spanish.

MEMORY - The Fyrite® INSIGHT has the ability to store, recall to dis- play, and print a minimum of 100 sets of time and date coded combustion, differential pressure, or differential temperature test records.

DOWNLOAD - Test records can be downloaded to a personal computer via the USB port. Saved records are transferred using PC based software and will be automatically loaded into an Excel spreadsheet. PC based software will also have the capability to update the instrument time and date and upload custom fuels, username, customer logo, and future code updates. The INSIGHT can also be set up from PC software.

USER INFORMATION - The instrument will accept up to three lines of 20 characters of user information. This information will appear with the test records when they are printed or downloaded.

CALIBRATION - Calibration can be performed by applying gas and enter- ing calibration gas concentrations through software menu selections.

SMART SENSORS - B-Smart sensor technology is employed for the CO channel. CO sensors are calibrated prior to shipment and provide data that can be entered through the software menu selections for easy calibration without applying gas.

CALIBRATION REMINDERS - Calibration reminders can be set for Never, 6, 8, 10, 12, or 15 months.

DISPLAY FORMAT - The Run/Hold display format can be presented in default or custom ordered for specific needs. Up to 8 lines can be custom- ized in the Run/Hold screen.

Fyrite® INSIGHT

DISPLAY ZOOM - Combustion test data in the Run/Hold screens can be shown with enlarged characters to make viewing easier. The complete list of Run/Hold data can be scrolled through for all zoom levels. Available zoom levels and information display are as follows:

Standard - 4 lines of information

2X - 3 lines of enlarged characters

3X - 2 lines of enlarged characters

STATUS - A status menu displays the instrument software version, boot version, model number, and serial number.

DIAGNOSTICS - The diagnostics menu displays the O2 sensor life, CO sensor information, and hours of operation for the instrument and the pump.

KEYPAD SOUND - Keypad sounds can be enabled or disabled with soft- ware menu selections.

LOW BATTERY - A tone will sound every 10 seconds when a low battery condition exists. A low battery message will also be displayed.

AUTOMATIC POWER DOWN - The instrument can be set to automati- cally power off after a period of inactivity unless the user presses any but- ton; the CO is greater than 50 ppm, or the O2 is less than 16.0%. Auto- matic power down can be set for None, 30, or 60 minutes.

PURGE - The sensors will be purged when the instrument is shutdown until sensor outputs indicate that they have been exposed to fresh air. Pressing the On/Off key a seconds time can defeat the purge routine. The minimum purge time can be set to None, 5 seconds, 1 , 5, or 10 minutes.

SENSOR ACCESS - Sensor replacement can be accomplished in the field. Easy access to the O2 and CO sensors is provided through the back of the instrument.

POWER - The Fyrite® INSIGHT is powered by 4 AA alkaline batteries. A new set of alkaline batteries will provide a minimum of 20 hours of con- tinuous operation with the pump running and the back light on. Option- ally, the analyzer can operate with a universal AC adapter. When the AC adapter is plugged into the unit, batteries do not have to be removed.

2

Instruction 24-9460

 

Instruction 24-9460

3

Image 4
Contents Fyrite Insight Table of Contents Introduction Fyrite InsightFyrite Insight Features & Benefits Technical Characteristics Operational OverviewSetting UP the Analyzer PowerConnecting the Probe Hose & Thermo- couple Preliminary StepsInsight Setup Mode Entering SetupTemperature Units Front Panel Push ButtonsPressure Units Clock5 O2 Reference Print PressureUsername 9 Run/Hold FormatButton Sound Language SelectionCAL Reminder Period Post-Purge Period Inactivity TimeoutDate Format Analyzer Power On & Warm Up OperationOperating Tips Fuel SelectionSampling Point Performing a Combustion TestMaking a Draft / Pressure Measurement Temperature Measurement Saving Test DataEnding a Combustion Test Emptying the Water TrapLow Battery Alarm Powering OFF the Analyzer & PurgingMemory Printing Test Data Fyrite User Software InstallationPress the Print F1 button to begin print- ing Windows Vista Operating System Calibration & Maintenance Fyrite User Software DescriptionSmart Sensors Starting a CalibrationPressure Sensor Calibration Materials RequiredTS-Zero Procedure ProcedureStack Calibration TS-Span Procedure TA-Zero ProcedureAir Calibration Material RequiredTA-Span Procedure CO Sensor CalibrationMaintenance Fyrite Insight Disassembly O2 Sensor Replacement Water Trap / FilterThermocouple Replacement CO Sensor ReplacementTools Required Gas Pump Replacement Cleaning the ProbeError Messages Parts & Tools RequiredDiagnostics Screen Status ScreenService Centers United States PennsylvaniaCanada MéxicoHeadquarters Hunt Valley Circle, New Kensington, PA

INSIGHT specifications

Bacharach INSIGHT is a cutting-edge platform designed for HVAC professionals and building management teams, providing real-time insights and analytics for optimal system performance. Offering a seamless integration of hardware and software, INSIGHT enhances operational efficiency by transforming traditional data management paradigms.

One of the standout features of Bacharach INSIGHT is its user-friendly interface, which enables users to visualize data in an easily digestible format. The platform presents key metrics, historical trends, and system alerts, empowering users to make informed decisions quickly. This accessibility is crucial for technicians and managers who need to identify potential issues before they escalate, ensuring systems run smoothly and effectively.

Another significant characteristic of Bacharach INSIGHT is its advanced data analytics capabilities. The platform leverages machine learning algorithms to analyze performance data, identify patterns, and predict maintenance needs. This proactive approach minimizes downtime and extends the lifespan of HVAC systems, resulting in cost savings for organizations.

INSIGHT also incorporates IoT technologies, allowing for real-time data collection from various connected devices across an HVAC system. This integration ensures that users have access to accurate and timely information, facilitating immediate responses to any irregularities. The use of cloud technology enables data storage and access from anywhere, promoting collaboration among team members and stakeholders.

The platform is designed with scalability in mind, accommodating small businesses with singular HVAC units as well as large enterprises with complex, multi-site systems. This flexibility means that as organizations grow, Bacharach INSIGHT can adapt to their changing needs without requiring extensive modifications.

Furthermore, user customization options allow for tailored dashboards and alerts, ensuring that each team can focus on the most relevant metrics for their operations. This adaptability contributes to improved user experience, enhancing system usability and engagement.

In summary, Bacharach INSIGHT combines intuitive design, powerful analytics, IoT integration, and scalability to redefine HVAC management. By equipping users with actionable insights and fostering proactive maintenance practices, INSIGHT positions organizations to optimize their energy use, reduce costs, and improve overall system reliability. This innovative platform is pioneering a new era for HVAC professionals, facilitating smarter building management with robust, real-time analytics.