Bacharach INSIGHT manual Air Calibration, TS-Span Procedure, Material Required, TA-Zero Procedure

Page 23

Fyrite® INSIGHT

2.If not already done, turn ON the analyzer and display the CALI- BRATION LIST screen per Section 5.2.

3.Use the ▲▼ buttons to highlight T-Stack,

and then press ENTER to display the CALI- BRATE TS-ZERO screen.

"Measured" is the current temperature read- ing, while "Applied" is a known temperature that will be applied for calibration purposes.

4.Set thermocouple simulator to 32 °F (0 °C), and then use the and buttons to enter an Applied value that exactly equals the setting of the simulator.

Alternatively: Submerge probe tip into an ice-water bath with a ther- mometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the thermometer read- ing.

The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to calibrate outside this range will cause the message “Applied Value High" (or Low) to appear at the bottom of the screen.

5. Wait until the Measured reading stabilizes, and then press ENTER to calibrate the TS-Zero Measured value to that of the Ap- plied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATE TS-SPAN screen.

TS-Span Procedure:

6.Set thermocouple simulator to 572 °F (300 °C), and then use the and buttons to enter an Applied value that exactly equals the setting of the simulator.

Alternatively: Submerge probe tip into a container of boiling water with a thermometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the ther- mometer reading.

Fyrite® INSIGHT

The calibration range is from 175 to 625 °F (80 to 330 °C). An at- tempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.

7.Wait until the Measured reading stabilizes, and then press ENTER to calibrate the TS-Span Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear fol- lowed by the CALIBRATION LIST screen being re-displayed.

5.6T-Air Calibration

This procedure first zeros and then spans the ambient-temperature chan- nel to known temperature values.

The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, ice and boil- ing water baths can be used.

Material Required:

Thermocouple simulator (K-type)

-Range: 0 to 600 °F

-Accuracy: ±0.5 °F

(Alternatively) Ice-water, boiling water, thermometer

TA-Zero Procedure:

1.Set thermocouple simulator to room temperature and plug its output into the T-AIR connector located at the bottom of the analyzer.

Alternatively: Plug the probe’s thermocouple into the T-AIR connector located at the bot- tom of the analyzer. DO NOT attach the probe’s gas hose to the analyzer’s GAS port; otherwise water will be drawn into the analyzer!

2.If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.

3.Use the buttons to highlight T-Air, and then press ENTER to display the CALI- BRATE TA-ZERO screen.

40

Instruction 24-9460

 

Instruction 24-9460

41

Image 23
Contents Fyrite Insight Table of Contents Fyrite Insight IntroductionFyrite Insight Features & Benefits Operational Overview Technical CharacteristicsPreliminary Steps Setting UP the AnalyzerPower Connecting the Probe Hose & Thermo- coupleInsight Front Panel Push Buttons Setup ModeEntering Setup Temperature UnitsPrint Pressure Pressure UnitsClock 5 O2 Reference9 Run/Hold Format UsernameCAL Reminder Period Language SelectionButton Sound Date Format Inactivity TimeoutPost-Purge Period Fuel Selection Analyzer Power On & Warm UpOperation Operating TipsPerforming a Combustion Test Sampling PointMaking a Draft / Pressure Measurement Emptying the Water Trap Temperature MeasurementSaving Test Data Ending a Combustion TestMemory Powering OFF the Analyzer & PurgingLow Battery Alarm Press the Print F1 button to begin print- ing Fyrite User Software InstallationPrinting Test Data Windows Vista Operating System Starting a Calibration Calibration & MaintenanceFyrite User Software Description Smart SensorsMaterials Required Pressure Sensor CalibrationStack Calibration ProcedureTS-Zero Procedure Material Required TS-Span ProcedureTA-Zero Procedure Air CalibrationCO Sensor Calibration TA-Span ProcedureMaintenance Fyrite Insight Disassembly Water Trap / Filter O2 Sensor ReplacementTools Required CO Sensor ReplacementThermocouple Replacement Parts & Tools Required Gas Pump ReplacementCleaning the Probe Error MessagesStatus Screen Diagnostics ScreenMéxico Service CentersUnited States Pennsylvania CanadaHeadquarters Hunt Valley Circle, New Kensington, PA

INSIGHT specifications

Bacharach INSIGHT is a cutting-edge platform designed for HVAC professionals and building management teams, providing real-time insights and analytics for optimal system performance. Offering a seamless integration of hardware and software, INSIGHT enhances operational efficiency by transforming traditional data management paradigms.

One of the standout features of Bacharach INSIGHT is its user-friendly interface, which enables users to visualize data in an easily digestible format. The platform presents key metrics, historical trends, and system alerts, empowering users to make informed decisions quickly. This accessibility is crucial for technicians and managers who need to identify potential issues before they escalate, ensuring systems run smoothly and effectively.

Another significant characteristic of Bacharach INSIGHT is its advanced data analytics capabilities. The platform leverages machine learning algorithms to analyze performance data, identify patterns, and predict maintenance needs. This proactive approach minimizes downtime and extends the lifespan of HVAC systems, resulting in cost savings for organizations.

INSIGHT also incorporates IoT technologies, allowing for real-time data collection from various connected devices across an HVAC system. This integration ensures that users have access to accurate and timely information, facilitating immediate responses to any irregularities. The use of cloud technology enables data storage and access from anywhere, promoting collaboration among team members and stakeholders.

The platform is designed with scalability in mind, accommodating small businesses with singular HVAC units as well as large enterprises with complex, multi-site systems. This flexibility means that as organizations grow, Bacharach INSIGHT can adapt to their changing needs without requiring extensive modifications.

Furthermore, user customization options allow for tailored dashboards and alerts, ensuring that each team can focus on the most relevant metrics for their operations. This adaptability contributes to improved user experience, enhancing system usability and engagement.

In summary, Bacharach INSIGHT combines intuitive design, powerful analytics, IoT integration, and scalability to redefine HVAC management. By equipping users with actionable insights and fostering proactive maintenance practices, INSIGHT positions organizations to optimize their energy use, reduce costs, and improve overall system reliability. This innovative platform is pioneering a new era for HVAC professionals, facilitating smarter building management with robust, real-time analytics.