3.0ALARMS AND INDICATORS _____________________________________

3.1TEMPERATURE PROBE ALARM

If the airway temperature probe is disconnected or develops a low or high resistance the Temperature Probe alarm will be enabled. Indication is by a ‘- - -' message on the airway temperature display plus a flashing Temperature Probe red LED and an audio alarm.

Five specific fault types can be diagnosed and are displayed by the sequence of ‘decimal points’ in the display, as shown below:

a

- - -

Temperature Probe not plugged in.Low temperature (~ 0 °C)

 

 

Open circuit chamber outlet probe

 

 

Open circuit in common lead in probe cable.

b- - -.Open circuit patient airway probe Very low temperature (~ 0 °C)

c

- -.-

Short circuit chamber outlet probe

 

 

Very high temperature (~ 100 °C)

d- -.-.Short circuit patient airway probe Very high temperature (~ 100 °C)

e

-.- -

Short in probe cable between patient airway probe and chamber outlet probe

The faults (a to d) above are checked for at a rate of about twice per second. The final fault (e) which does not immediately cause any apparent operating anomaly is checked twice a minute.

A Temperature Probe Alarm immediately shuts down both heater wire and heaterplate.

Once the alarm condition has been removed the humidifier will enter warm up mode (refer § 2.6).

3.2ALARMS IN HEATER WIRE MODE (MR700/720/730 ONLY)

3.2.1HEATER WIRE ALARM

If the heater wire is disconnected or there is a fault with the heater wire circuit, then the Heater Wire alarm LED will start flashing and the audio alarm will be enabled. Both heater wire and heaterplate circuits will be shutdown. The display will continue to indicate the temperature sensor reading (unless the heater wire fuse is open circuit).

An open circuit heater wire fuse is indicated by ‘F-2’ appearing on the Airway Temperature display.

On MR720/730 if the Heater Wire is connected but switched off, the alarm will sound (refer § 2.2.4).

Note: If ‘F-2’ is displayed and the F2 fuse is intact then check the calibration of the 41 °C overheat (refer § 5).

3.2.2AIRWAY TEMPERATURE HIGH ALARM

The red Airway Temp High LED will flash and an audible alarm will sound if the gas temperature being delivered to the patient is 2 °C higher than the Temperature Control setting. A high temperature alarm instantly disables both heater circuits.

3.2.3AIRWAY TEMPERATURE LOW ALARM

The red Airway Temp Low LED will flash and an audible alarm will sound if the gas temperature being delivered to the patient is 2 °C lower (4 °C for MR720 model) than the Temperature Control setting. If Airway Temperature Low alarm condition persists for more than 100 seconds the heaterplate and heater wire circuits will switch off. Pressing the mute button will silence the Audio alarm and resume heat for a further 100 seconds, the Airway Temperature Low LED will continue to flash. This alarm is disabled during warm-up. See § 2.6.1.

If the airway temperature drops below 5°C the display will show ‘LO’ and an audible alarm will sound.

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Image 12
Contents D E L S MR730 MR720 MR700 MR480 Auckland, New ZealandRevision F changes Section Description of Change National OfficeContents Dual Probe Temperature Test Performance Check MR480Power PCB Harness Connection Diagram Maintenance Schedule B1.1Maintenance Schedule Checklist 10.1Appendix D MR700 Series General Specifications Models MR730/720/700EOH SETTOP RSDOperating Controls and Modes MR700 Series Humidifiers with Version 1.0 SoftwareStandby Button MR700/720/730 only 2 MR720 Version 2.0, 3.0 Software 1 MR700/MR730 ALL Software VERSIONS, MR720 Version 1.0 onlyAlarms and Indicators Version 1.0 and Version 2.0 Software Version 3.0 SoftwareAudio Alarm Circuit Operation Power Line Interrupt Revisions D/F Pcbs Revision H and J Pcbs Calibration For models with Revision D PCB and earlierPage Performance Check Alarms Check MR700/720/730 Dual Probe Temperature Test Replacing revision F or earlier power PCB MR700/720/730 Servicing InformationReplacing revision F or earlier power PCB MR480 Fault Location SCHEMATICS, Part Lists and Mechanical Drawings Power Board Control Board Component Layout Revision F PCB Page This page Left Blank Circuit Diagrams Revision H and J PCB Control PCB Circuit Diagrams Revision H and J PCB Power PCB Circuit Diagrams Revision F PCB Control PCB Circuit Diagrams Revision F PCB Power PCB DSP1, DSP2, DSP3 LED1, LED3LED2,LED4, LED5,LED11 LED6,LED7 LED8,LED9,LED10SW5 SPK1VR1 VR4Fuse fast acting 2 a 20 × 5 mm for 230 999830009 MR480AGM Exploded Diagrams MR480 10.3 MR730/720/700 Mechanical Parts MR720ASU Exploded Diagrams MR730/720/700 Heaterplate Assembly Parts MR730/720/700/480 Page PCB Identification Diagram Power PCB Harness Connection Diagram Product Change History Date Change First Serial Number Comments AffectedAAA Xxxxx Year Model Code Serial NumberAppendix a OFFError Messages A2.3 Integral DisplayEP1 EP2EP3 EP4First Byte Second ByteThird Byte A5.4 Serial Data Interface Connector B1 Maintenance Schedule Appendix BPhotocopy and File in Maintenance Records Electrical Safety Checks as requiredAppendix C Possible cause Action to takeHigh Temperature C1.5 C1.6LED on Check Voltage Rating Of Unit Cleaning instructions Appendix DSetup Notes Appendix EDynamic Test Sequence MR730/720/700 Dynamic Test Procedure MR480