Fisher & Paykel MR480, MR720, MR730, MR700 technical manual Audio Alarm

Page 14

8

Pushing the mute button mutes the audio alarm for ten minutes and restores power to the heaterplate during this time. The Airway Temp Low LED will continue to flash. This alarm is disabled during warmup. See § 2.6.2.

If the airway temperature drops below 5°C the display will show ‘LO’ and an audible alarm will sound.

3.4AUDIO ALARM

A choice of the ISO draft standard audio alarm or a gliding tone can be selected by the programming control link (L1). Refer Appendix A3.

A momentary depression of the mute button will silence the audio alarm for about 3 minutes (except for a low temperature alarm in no heater wire mode. See § 3.3.3. The relevant LED indicators remain flashing until the alarm condition clears. Any new alarm condition occurring within a mute delay will cause the audio alarm to be re-enabled.

3.5SAFETY PROTECTION (ALL MODELS)

3.5.1TEMPERATURE DEVIATIONS

An over temperature alarm condition causes an immediate shut down of both heater circuits. In the case of a Low Temperature alarm, pressing Mute causes heating to recommence, the Low Temperature LED however remains on until the fault has cleared. There are two independent backup over temperature protection circuits.

Firstly, an over temperature protection circuit independent of the microprocessor control continuously monitors the airway thermistor sensor. This backup protection is preset to disconnect the heating circuits at 41.0 °C.Secondly, a bimetallic thermostat fixed to the heaterplate operates if the heaterplate temperature exceeds 118 ± 6 °C and cuts out the mains active supply. On cooling, this protector can only be reset by gaining access to the under surface of the heaterplate and manually depressing the red reset button.

Heaterplate temperature is limited to 110 °C max during normal operation (80 °C for MR720). Optional heaterplate limits of 50, 80 and 90 °C can be selected. (Refer Appendix A3).

3.5.2MICROPROCESSOR WATCH DOG

A control loop in the operating software continuously monitors for correct operation. Should a fault be detected a hardware reset circuit attempts to reinitialise the microprocessor. If this is not successful additional hardware disables the heating circuit, blanks the display and turns the Humidifier Fault red LED on.

The watch dog circuit is tested at turn on by deliberately halting the processor and forcing initialisation.

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Image 14
Contents D E L S MR730 MR720 MR700 MR480 Auckland, New ZealandRevision F changes Section Description of Change National OfficeContents Dual Probe Temperature Test Performance Check MR480Maintenance Schedule Checklist Power PCB Harness Connection DiagramMaintenance Schedule B1.1 10.1Appendix D MR700 Series General Specifications Models MR730/720/700TOP EOHSET RSDOperating Controls and Modes MR700 Series Humidifiers with Version 1.0 SoftwareStandby Button MR700/720/730 only 2 MR720 Version 2.0, 3.0 Software 1 MR700/MR730 ALL Software VERSIONS, MR720 Version 1.0 onlyAlarms and Indicators Version 1.0 and Version 2.0 Software Version 3.0 SoftwareAudio Alarm Circuit Operation Power Line Interrupt Revisions D/F Pcbs Revision H and J Pcbs Calibration For models with Revision D PCB and earlierPage Performance Check Alarms Check MR700/720/730 Dual Probe Temperature Test Replacing revision F or earlier power PCB MR700/720/730 Servicing InformationReplacing revision F or earlier power PCB MR480 Fault Location SCHEMATICS, Part Lists and Mechanical Drawings Power Board Control Board Component Layout Revision F PCB Page This page Left Blank Circuit Diagrams Revision H and J PCB Control PCB Circuit Diagrams Revision H and J PCB Power PCB Circuit Diagrams Revision F PCB Control PCB Circuit Diagrams Revision F PCB Power PCB LED2,LED4, LED5,LED11 DSP1, DSP2, DSP3LED1, LED3 LED6,LED7 LED8,LED9,LED10VR1 SW5SPK1 VR4Fuse fast acting 2 a 20 × 5 mm for 230 999830009 MR480AGM Exploded Diagrams MR480 10.3 MR730/720/700 Mechanical Parts MR720ASU Exploded Diagrams MR730/720/700 Heaterplate Assembly Parts MR730/720/700/480 Page PCB Identification Diagram Power PCB Harness Connection Diagram Product Change History Date Change First Serial Number Comments AffectedAAA Xxxxx Year Model Code Serial NumberAppendix a OFFError Messages A2.3 Integral DisplayEP3 EP1EP2 EP4Third Byte First ByteSecond Byte A5.4 Serial Data Interface Connector B1 Maintenance Schedule Appendix BPhotocopy and File in Maintenance Records Electrical Safety Checks as requiredAppendix C Possible cause Action to takeHigh Temperature C1.5 C1.6LED on Check Voltage Rating Of Unit Cleaning instructions Appendix DSetup Notes Appendix EDynamic Test Sequence MR730/720/700 Dynamic Test Procedure MR480