Fisher & Paykel MR700, MR720, MR730, MR480 technical manual EP1, EP2, EP3, EP4, A5.1

Page 51

 

 

 

45

 

Errors EP2 to EP4 can be caused for example by abnormal high differential temperatures on the two thermistors or by

 

rapidly changing input temperatures.

 

EP1

Hardware fault; Q13 short not effective.

 

EP2

Low probe temperature (~0 °C).

 

EP3

High probe temperature (~70 °C).

 

EP4

Partial short between the active probe cable conductors.

A5

MR700 SERIES HUMIDIFIER SERIAL DATA INTERFACE

A5.1

SERIAL DATA FORMAT

 

1 byte

 

: (semi colon) Marks beginning of data record.

 

4 bytes ASCII

Temperature Control setting.

 

4 bytes ASCII

Airway temperature.

 

4 bytes ASCII

Chamber temperature.

 

4 bytes ASCII

Heaterplate temperature.

 

4 bytes ASCII

Heater wire duty cycle.

 

4 bytes ASCII

Heaterplate duty cycle.

 

4 bytes ASCII

Chamber Control setting .

 

4 bytes *

 

Status information. (1*ASCII 3*binary bytes).

 

2 bytes

 

CR, LF (Marks end of data record).

Between each of the above lines there is approximately a 50 ms pause before the next line is transmitted. The complete data record is repeated every 500 to 600 ms (500 ms on a 60 Hz supply, 600 ms with 50 Hz).

Data format is RS232 levels, no handshaking, 9600 baud, 8 bits, no parity, with one stop bit.

Between the semi colon and the line feed there are eight, 4-byte packets of information as shown above.

In each case the first byte of the packet contains an identification letter from ‘a’ through ‘h’ in ASCII.

The format of the 3 bytes following the ASCII I.D. is as shown below.

I.D. a - SET temperature.

I.D. b - Airway temperature.

I.D. c - Chamber temperature.

Range 5.0 to 80.0 °C, decimal point is assumed between 1st and 2nd digits.

I.D. d - Heaterplate temperature.

Range 14.0 to 99.9 °C decimal point assumed, 100 to 135°C no decimal point.

I.D. e - Heater Wire duty cycle.

I.D. f - Heaterplate duty cycle.

Range 0 to 100 %.

I.D. g - Chamber Control Setting.

Range -5.0 to +2.0 decimal point assumed. Plus or minus sign in leading ASCII byte.

I.D. h - Status information.

A5.2 STATUS INFORMATION

The ASCII I.D. is followed by three binary bytes giving alarms and error information, see list below:

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Image 51
Contents Auckland, New Zealand D E L S MR730 MR720 MR700 MR480National Office Revision F changes Section Description of ChangeContents Performance Check MR480 Dual Probe Temperature Test10.1 Power PCB Harness Connection DiagramMaintenance Schedule B1.1 Maintenance Schedule ChecklistAppendix D Models MR730/720/700 MR700 Series General SpecificationsRSD EOHSET TOPMR700 Series Humidifiers with Version 1.0 Software Operating Controls and ModesStandby Button MR700/720/730 only 1 MR700/MR730 ALL Software VERSIONS, MR720 Version 1.0 only 2 MR720 Version 2.0, 3.0 SoftwareAlarms and Indicators Version 3.0 Software Version 1.0 and Version 2.0 SoftwareAudio Alarm Circuit Operation Power Line Interrupt Revisions D/F Pcbs Revision H and J Pcbs For models with Revision D PCB and earlier CalibrationPage Performance Check Alarms Check MR700/720/730 Dual Probe Temperature Test Servicing Information Replacing revision F or earlier power PCB MR700/720/730Replacing revision F or earlier power PCB MR480 Fault Location SCHEMATICS, Part Lists and Mechanical Drawings Power Board Control Board Component Layout Revision F PCB Page This page Left Blank Circuit Diagrams Revision H and J PCB Control PCB Circuit Diagrams Revision H and J PCB Power PCB Circuit Diagrams Revision F PCB Control PCB Circuit Diagrams Revision F PCB Power PCB LED6,LED7 LED8,LED9,LED10 DSP1, DSP2, DSP3LED1, LED3 LED2,LED4, LED5,LED11VR4 SW5SPK1 VR1Fuse fast acting 2 a 20 × 5 mm for 230 999830009 MR480AGM Exploded Diagrams MR480 10.3 MR730/720/700 Mechanical Parts MR720ASU Exploded Diagrams MR730/720/700 Heaterplate Assembly Parts MR730/720/700/480 Page PCB Identification Diagram Power PCB Harness Connection Diagram Date Change First Serial Number Comments Affected Product Change HistoryYear Model Code Serial Number AAA XxxxxOFF Appendix aA2.3 Integral Display Error MessagesEP4 EP1EP2 EP3First Byte Second ByteThird Byte A5.4 Serial Data Interface Connector Appendix B B1 Maintenance ScheduleElectrical Safety Checks as required Photocopy and File in Maintenance RecordsPossible cause Action to take Appendix CHigh Temperature C1.6 C1.5LED on Check Voltage Rating Of Unit Appendix D Cleaning instructionsAppendix E Setup NotesDynamic Test Sequence MR730/720/700 Dynamic Test Procedure MR480