Fisher & Paykel MR720, MR730, MR700 technical manual Replacing revision F or earlier power PCB MR480

Page 24

18

If you are fitting a Revision H or later power PCB to a Humidifier that has a revision F or earlier control PCB , you will have to ensure correct phase polarity on the transformer secondary wires. To check for correct polarity, turn the humidifier on, if the humidifier displays F-2, then the polarity of one of the transformer secondaries needs to be changed.

This is accomplished as follows:

i.Unplug the power supply and then unplug the transformer secondary connector (Connector J9) from the power PCB.

ii.Then carefully remove the two yellow wires from the connector housing by depressing the locking tabs on the back of the connector terminals and pulling the wires out of the housing.

iii.Insert the yellow wires in to the connector housing in reverse order to what they were before. See below.

YELLOW

YELLOW

WHITE

WHITE

J9

YELLOW

YELLOW

WHITE

WHITE

J9

SECONDARY EXISTING WIRING

SECONDARY

NEW WIRING AFTER

POLARITY SWAPPED

Replacing revision F or earlier power PCB MR480

If the existing power PCB is Rev. F or earlier, there will be an earth wire from the secondary harness of the transformer to the chassis. Unclip this wire from the transformer connector and unscrew it from the chassis. The Rev. J replacement power PCB will have an earth wire attached to the PCB. Fit the end of this earth wire to the earth stud on the chassis.

7.3.2MR480 TRANSFORMERS

For MR480’s , the transformer secondary wiring changed when Rev. J PCB started. When replacing the transformer, follow the instructions supplied in the transformer replacement kits to ensure compatability.

7.3.3SOFTWARE COMPATABILITY

Software versions 1.0 and 2.0 can be upgraded to version 3.0 in their existing PCBs.

Software version 1.0 should be replaced with version 3.0 if the PCB’s are upgraded. Software versions 2.0 and 3.0 are compatable with all PCB revisions.

7.4HEATERPLATE

To reset the 118 °C over temperature cutout, separate the case front from the case body, then push the reset button (through a small hole under the heaterplate in the front panel case assembly).

To replace the heaterplate:

1Separate the case front from the humidifier body.

2Unclip the Power PCB from the case body (see § 7.2.2 above).

3Unplug the heaterplate wires from the Power PCB (connectors J3 and J7).

4Unscrew the earth screw on the chassis for the heaterplate earth wire.

5Unplug the ribbon cables from the Power PCB to the Control PCB (connectors J1 and J2). The front case assembly is now separate from the case body.

6Detach the heaterplate from the case front by unscrewing the three screws which secure it to the case front.

7Assembly is the reverse of the above procedure.

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Image 24
Contents D E L S MR730 MR720 MR700 MR480 Auckland, New ZealandRevision F changes Section Description of Change National OfficeContents Dual Probe Temperature Test Performance Check MR480Power PCB Harness Connection Diagram Maintenance Schedule B1.1Maintenance Schedule Checklist 10.1Appendix D MR700 Series General Specifications Models MR730/720/700EOH SETTOP RSDOperating Controls and Modes MR700 Series Humidifiers with Version 1.0 SoftwareStandby Button MR700/720/730 only 2 MR720 Version 2.0, 3.0 Software 1 MR700/MR730 ALL Software VERSIONS, MR720 Version 1.0 onlyAlarms and Indicators Version 1.0 and Version 2.0 Software Version 3.0 SoftwareAudio Alarm Circuit Operation Power Line Interrupt Revisions D/F Pcbs Revision H and J Pcbs Calibration For models with Revision D PCB and earlierPage Performance Check Alarms Check MR700/720/730 Dual Probe Temperature Test Replacing revision F or earlier power PCB MR700/720/730 Servicing InformationReplacing revision F or earlier power PCB MR480 Fault Location SCHEMATICS, Part Lists and Mechanical Drawings Power Board Control Board Component Layout Revision F PCB Page This page Left Blank Circuit Diagrams Revision H and J PCB Control PCB Circuit Diagrams Revision H and J PCB Power PCB Circuit Diagrams Revision F PCB Control PCB Circuit Diagrams Revision F PCB Power PCB DSP1, DSP2, DSP3 LED1, LED3LED2,LED4, LED5,LED11 LED6,LED7 LED8,LED9,LED10SW5 SPK1VR1 VR4Fuse fast acting 2 a 20 × 5 mm for 230 999830009 MR480AGM Exploded Diagrams MR480 10.3 MR730/720/700 Mechanical Parts MR720ASU Exploded Diagrams MR730/720/700 Heaterplate Assembly Parts MR730/720/700/480 Page PCB Identification Diagram Power PCB Harness Connection Diagram Product Change History Date Change First Serial Number Comments AffectedAAA Xxxxx Year Model Code Serial NumberAppendix a OFFError Messages A2.3 Integral DisplayEP1 EP2EP3 EP4First Byte Second ByteThird Byte A5.4 Serial Data Interface Connector B1 Maintenance Schedule Appendix BPhotocopy and File in Maintenance Records Electrical Safety Checks as requiredAppendix C Possible cause Action to takeHigh Temperature C1.5 C1.6LED on Check Voltage Rating Of Unit Cleaning instructions Appendix DSetup Notes Appendix EDynamic Test Sequence MR730/720/700 Dynamic Test Procedure MR480