|
| 51 |
|
C.1.2 | HIGH TEMPERATURE |
|
|
|
|
|
|
| Condition detected | Possible cause | F Action to take |
|
|
|
|
|
|
|
|
| Airway temperature high alarm | Sudden spurious increase in gas flow. | Wait for temperature to fall below |
| activated. |
| threshold. |
|
|
|
|
|
| Heater wire too close to temperature | Use shorter heater wire or lengthen |
|
| probe. | inspiratory hose so that distance is |
|
|
| between 25 (1”) and 100 mm (4”), ideally |
|
|
| 30 mm. |
|
|
|
|
|
| Possibly faulty EOH sensor on | Check probe, see § 6.4. |
|
| temperature probe (low resistance). |
|
|
|
|
|
| Chamber temperature high alarm | No water in chamber. | Refill humidification chamber. |
| activated. |
|
|
|
|
|
|
|
| Possibly faulty TOP sensor on | Check probe, see § 6.4. |
|
| temperature probe (low resistance). |
|
|
|
|
|
C.1.3 | TEMPERATURE FLUCTUATING |
|
|
|
|
|
|
| Condition detected | Possible cause | FAction to take |
|
|
|
|
|
|
|
|
| High and low temperature | Ventilation rate too low or too erratic | Ventilation rates below about 6 bpm or of |
| alarms. | for proper humidifier temperature | a highly erratic nature may affect the |
|
| control. | stability of the temperature control. |
|
|
|
|
|
| Humidifier temperature control | Shield breathing circuit from external |
|
| influenced by external heating or | influence. |
|
| cooling such as air conditioner. |
|
|
|
|
|
|
| Temperature probes incorrectly | Check temperature probe position in the |
|
| positioned. | circuit. |
|
|
|
|
|
| Possibly faulty temperature probe | Check probe, see Section 6.4. |
|
| (Could be intermittent short or open | (Wriggle probe to pick up intermittent |
|
| circuit). | |
|
| faults). | |
|
|
| |
|
|
|
|
C.1.4 | EXCESSIVE WATER IN CIRCUIT |
|
|
|
|
|
|
| Condition detected | Possible cause | FAction to take |
|
|
|
|
|
|
|
|
| Excessive water in inspiratory | If in | Reposition water trap. |
| circuit due to rainout. | not at the lowest point of the circuit. |
|
|
|
|
|
|
| If in heated wire mode, heater wire too | Increase distance from heater wire to |
|
| close to temperature probe. | temperature probe so that the distance is |
|
|
| between 25 (1”) and 100mm (4”), ideally |
|
|
| 30mm. |
|
|
|
|
|
| Inspiratory circuit being cooled by air | Shield circuit from air conditioner or use |
|
| conditioner. | heater wire circuits. |
|
|
|
|
|
| Chamber control set too high | Reduce setting. |
|
|
|
|
|
| Maximum permissible peak flow rate of | Use correct chamber. Refer to the |
|
| humidification chamber has been | operating instructions for the type of |
|
| exceeded. | chamber used. |
|
|
|
|
|
| Humidifier chamber tilted. | Ensure chamber is level. |
|
|
|
|
|
| Possibly faulty temperature probe (low | Check probe, see § 6.4 |
|
| resistance). |
|
|
|
|
|
|
| Inspiratory heater wire not connected | Connect inspiratory heater wire. |
|
| in a parallel wired dual circuit. |
|
|
|
|
|
Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001