Fisher & Paykel MR720, MR730, MR480, MR700 technical manual Check Voltage Rating Of Unit

Page 60

54

C2 MR700 SERIES FAULT FLOW CHART FOR NO GO CONDITION

Check Voltage

Rating Of Unit

YES

Open Unit and check if thermal cut out under heaterplate has tripped

NO

Check for mains voltage at Connector J5 on power board

YES

Reset thermal cut out and find reason for fault

NO

Check mains lead and plug

YES

Check connectors on power board are in correct locations and secure, check fuses (F1 - F4), check voltage on connector J7of power PCB, mains switch on, connectors on rear of control board and continuity of transformer, connector J8-1& 3 (Primary) and connector J9-1 & 2

+3 & 4 (Secondaries).Check voltage on transformer

secondary J9.

BAD

Repair faulty component and test

GOOD

Replace power board and test

NO

Replace control board and test

NO

Return unit for repair

YES

Assemble and test

YES

Perform calibration and performance checks as outlined in Sections 3.0 and

4.0of the Technical Manual before re-instatement of unit

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Image 60
Contents D E L S MR730 MR720 MR700 MR480 Auckland, New ZealandRevision F changes Section Description of Change National OfficeContents Dual Probe Temperature Test Performance Check MR480Power PCB Harness Connection Diagram Maintenance Schedule B1.1Maintenance Schedule Checklist 10.1Appendix D MR700 Series General Specifications Models MR730/720/700EOH SETTOP RSDOperating Controls and Modes MR700 Series Humidifiers with Version 1.0 SoftwareStandby Button MR700/720/730 only 2 MR720 Version 2.0, 3.0 Software 1 MR700/MR730 ALL Software VERSIONS, MR720 Version 1.0 onlyAlarms and Indicators Version 1.0 and Version 2.0 Software Version 3.0 SoftwareAudio Alarm Circuit Operation Power Line Interrupt Revisions D/F Pcbs Revision H and J Pcbs Calibration For models with Revision D PCB and earlierPage Performance Check Alarms Check MR700/720/730 Dual Probe Temperature Test Replacing revision F or earlier power PCB MR700/720/730 Servicing InformationReplacing revision F or earlier power PCB MR480 Fault Location SCHEMATICS, Part Lists and Mechanical Drawings Power Board Control Board Component Layout Revision F PCB Page This page Left Blank Circuit Diagrams Revision H and J PCB Control PCB Circuit Diagrams Revision H and J PCB Power PCB Circuit Diagrams Revision F PCB Control PCB Circuit Diagrams Revision F PCB Power PCB DSP1, DSP2, DSP3 LED1, LED3LED2,LED4, LED5,LED11 LED6,LED7 LED8,LED9,LED10SW5 SPK1VR1 VR4Fuse fast acting 2 a 20 × 5 mm for 230 999830009 MR480AGM Exploded Diagrams MR480 10.3 MR730/720/700 Mechanical Parts MR720ASU Exploded Diagrams MR730/720/700 Heaterplate Assembly Parts MR730/720/700/480 Page PCB Identification Diagram Power PCB Harness Connection Diagram Product Change History Date Change First Serial Number Comments AffectedAAA Xxxxx Year Model Code Serial NumberAppendix a OFFError Messages A2.3 Integral DisplayEP1 EP2EP3 EP4First Byte Second ByteThird Byte A5.4 Serial Data Interface Connector B1 Maintenance Schedule Appendix BPhotocopy and File in Maintenance Records Electrical Safety Checks as requiredAppendix C Possible cause Action to takeHigh Temperature C1.5 C1.6LED on Check Voltage Rating Of Unit Cleaning instructions Appendix DSetup Notes Appendix EDynamic Test Sequence MR730/720/700 Dynamic Test Procedure MR480