Fisher & Paykel MR720, MR730, MR480, MR700 technical manual Appendix C, Possible cause Action to take

Page 56

50

APPENDIX C __________________________________________________

C1 TROUBLESHOOTING GUIDE FOR THE MR700 SERIES RESPIRATORY HUMIDIFIERS

This is a guide to possible corrective action in the case of an humidifier alarm condition or malfunction. It is primarily written for humidifiers using heated wire mode. So parts of this guide would not be applicable for the MR480 and MR730 and MR720 in the heater wire off mode.

Unless the cause and immediate effects of an alarm condition are understood and assessed to be of no hazard to the patient, the humidifier should be immediately switched off and removed from the patient circuit.

C1.1 LOW TEMPERATURE

Condition detected

Possible cause

F Action to take

 

 

 

 

 

 

Temperature less than desired,

Humidifier warming up.

Wait for a few minutes.

no alarm.

 

 

 

 

 

 

Standby mode switched on.

Push mode button for 1 second to

 

 

disable standby. Wait 20 minutes (30

 

 

minutes for v3.0 software).

 

 

 

 

Temperature probe not in position.

Check position of temperature probe in

 

 

the circuit.

 

 

 

Airway temperature low alarm

Gas flow stopped or interrupted.

Enter standby mode or turn humidifier

activated.

 

off until gas flow resumed. (Gas flow

 

 

must be resumed before coming out of

 

 

standby mode).

 

 

 

 

Set temperature increased by more than

Wait for temperature to rise.

 

2 °C.

 

 

 

 

 

Temperature probe not in probe

Fit temperature probe.

 

housing.

 

 

 

 

 

Gas flow rate too low.

Increase gas flow to more than 2 litres

 

 

per minute.

 

 

 

 

Too much distance between end of

Use longer heater wire or shorten

 

heater wire and temperature probe.

inspiratory hose so that distance is

 

 

between 25 (1") and 100 mm (4"), ideally

 

 

30 mm.

 

 

 

Chamber temperature low alarm

Poor thermal contact between base of

Check that chamber base and heaterplate

activated.

chamber and heaterplate.

are clean, smooth and flat. If necessary

 

 

clean or replace. Humidification

 

 

chambers which are not manufactured by

 

 

Fisher & Paykel may not be suitable.

 

 

 

 

Temperature probe not inserted at

Fit temperature probe correctly.

 

chamber outlet.

 

 

 

 

 

Heater element needs replacement.

Send for repair.

 

 

 

 

Gas flow rate too high for chamber

Use correct chamber. Refer to the

 

 

operating instructions for the type of

 

 

chamber used.

 

 

 

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Image 56
Contents D E L S MR730 MR720 MR700 MR480 Auckland, New ZealandRevision F changes Section Description of Change National OfficeContents Dual Probe Temperature Test Performance Check MR480Power PCB Harness Connection Diagram Maintenance Schedule B1.1Maintenance Schedule Checklist 10.1Appendix D MR700 Series General Specifications Models MR730/720/700EOH SETTOP RSDOperating Controls and Modes MR700 Series Humidifiers with Version 1.0 SoftwareStandby Button MR700/720/730 only 2 MR720 Version 2.0, 3.0 Software 1 MR700/MR730 ALL Software VERSIONS, MR720 Version 1.0 onlyAlarms and Indicators Version 1.0 and Version 2.0 Software Version 3.0 SoftwareAudio Alarm Circuit Operation Power Line Interrupt Revisions D/F Pcbs Revision H and J Pcbs Calibration For models with Revision D PCB and earlierPage Performance Check Alarms Check MR700/720/730 Dual Probe Temperature Test Replacing revision F or earlier power PCB MR700/720/730 Servicing InformationReplacing revision F or earlier power PCB MR480 Fault Location SCHEMATICS, Part Lists and Mechanical Drawings Power Board Control Board Component Layout Revision F PCB Page This page Left Blank Circuit Diagrams Revision H and J PCB Control PCB Circuit Diagrams Revision H and J PCB Power PCB Circuit Diagrams Revision F PCB Control PCB Circuit Diagrams Revision F PCB Power PCB DSP1, DSP2, DSP3 LED1, LED3LED2,LED4, LED5,LED11 LED6,LED7 LED8,LED9,LED10SW5 SPK1VR1 VR4Fuse fast acting 2 a 20 × 5 mm for 230 999830009 MR480AGM Exploded Diagrams MR480 10.3 MR730/720/700 Mechanical Parts MR720ASU Exploded Diagrams MR730/720/700 Heaterplate Assembly Parts MR730/720/700/480 Page PCB Identification Diagram Power PCB Harness Connection Diagram Product Change History Date Change First Serial Number Comments AffectedAAA Xxxxx Year Model Code Serial NumberAppendix a OFFError Messages A2.3 Integral DisplayEP1 EP2EP3 EP4Third Byte First ByteSecond Byte A5.4 Serial Data Interface Connector B1 Maintenance Schedule Appendix BPhotocopy and File in Maintenance Records Electrical Safety Checks as requiredAppendix C Possible cause Action to takeHigh Temperature C1.5 C1.6LED on Check Voltage Rating Of Unit Cleaning instructions Appendix DSetup Notes Appendix EDynamic Test Sequence MR730/720/700 Dynamic Test Procedure MR480