Fluke 744 manual Hart Mode Menu Tree

Page 17

Documenting Process Calibrator HART Mode

Communication Operations

PROCESS

SETUP

SERVICE

ABORT

(Example)

PV

%Range

AO

SV, TV, QV

Loop test

Pressure zero trim

Output trim

Sensor trim

Basic

Sensor

Device Identification

HART

Output

HART Information

Tag

PV units

LRV, URV

Damping

Transfer function

Software version

Final assembly number

Tag

Message

Date

Description

Manufacturer

Model

Device HART ID

Software revision

Hardware revision

Number of preambles

Sensor serial number

Sensor lower and upper limits

Sensor minimum span

(temperature devices only)

Sensor type

Sensor connections

Write protect

Alarm state

HART poll address

HART burst mode

HART burst command

pe03f.eps

Figure 2. 744 HART Mode Menu Tree

9

Image 17
Contents 744 Page Table of Contents 744 Hart Mode Users GuideList of Tables 744 List of Figures 744 Introduction Documenting Process Calibrator Hart ModeHow to Contact Fluke Connecting to a Hart Transmitter ∙ Poll address if not Documenting Process Calibrator Hart Mode Supported vs. Generic Transmitters Supported vs. Generic Transmitters Viewing Process Variables Setup OperationsCommunication Operations Hart Mode Menu Tree Sensor Hart Output Hart Information Abort Softkey Interaction between Analog Mode and Hart ModeService Operations Understanding Hart Calibration Hart Mode Menus for AdjustmentBlock Diagram of a Hart Transmitter Calibrating a Supported Hart Transmitter Loop Test Sensor Trim Output TrimPress Continue Cloning a Transmitter 744 Index Device identification